Chinese Journal of Contemporary Pediatrics ; (12): 809-813, 2021.
Artículo en Inglés | WPRIM | ID: wpr-888486

RESUMEN

OBJECTIVES@#To evaluate the accuracy and safety of measurements of transcutaneous carbon dioxide partial pressure (TcPCO@*METHODS@#A total of 45 very low birth weight infants were enrolled. TcPCO@*RESULTS@#There was no significant difference in TcPCO@*CONCLUSIONS@#Lower electrode temperatures (38-41℃) can accurately measure blood carbon dioxide partial pressure in very low birth weight infants, and thus can be used to replace the electrode temperature of 42°C. Transcutaneous measurements at the lower electrode temperatures may be helpful for understanding the changing trend of blood oxygen partial pressure.


Asunto(s)
Humanos , Lactante , Recién Nacido , Monitoreo de Gas Sanguíneo Transcutáneo , Dióxido de Carbono , Electrodos , Recién Nacido de muy Bajo Peso , Oxígeno , Presión Parcial , Temperatura