Journal of Xi'an Jiaotong University(Medical Sciences) ; (6): 977-982, 2023.
Artículo en Chino | WPRIM | ID: wpr-1005785

RESUMEN

【Objective】 To evaluate the effect of one-beat acquisition with wide detector CT on the image quality and diagnostic efficiency of coronary CT angiography (CCTA) in patients with atrial fibrillation. 【Methods】 A total of 52 consecutive patients with atrial fibrillation, including 31 males, (67.32±11.45) years old, who underwent CCTA from July 2022 to February 2023, were analyzed retrospectively. All patients underwent one-beat acquisition CCTA. The subjective and objective image quality of the coronary arteries was evaluated, and using invasive coronary catheter angiography as the gold standard, the diagnostic efficacy of stenosis degrees above moderate and severe degrees was calculated, respectively. 【Results】 Subjective evaluation results: 92.31% (384/416) of the vascular segments were rated as excellent or good, and the diagnosable rate reached 98.08% (408/416, subjective score ≥3 points). Objective evaluation results: The CT value of the right coronary artery, anterior descending branch, and circumflex branch was (433.41±95.17)HU, (422.69±92.81)HU and (420.27±95.43)HU, respectively; the contrast-to-noise ratio was 38.46±7.54, 32.46±13.78 and 37.74±8.89, respectively. The total diagnostic accuracy, sensitivity, and specificity was 94.71%, 87.9% and 96.62%, respectively, for moderate stenosis and 96.15%, 83.64% and 98.06% for severe stenosis. 【Conclusion】 One-beat acquisition with wide detector CT can obtain high-quality coronary artery images and high diagnostic accuracy for patients with atrial fibrillation without radiation dose increase to patients. It has good clinical application value for patients with atrial fibrillation.

2.
Journal of Xi'an Jiaotong University(Medical Sciences) ; (6): 905-911, 2022.
Artículo en Chino | WPRIM | ID: wpr-1006646

RESUMEN

【Objective】 To investigate the improvement of signal uniformity in cardiac magnetic resonance image by adjusting the transmit gain (TG). 【Methods】 For this study we recruited 11 volunteers for cardiac MR scans between July and September 2018. The scanner of US GE Discovery 750 3.0T superconducting MRI was used. We divided the volunteers into normal group and overweight group according to their individual BMI index. The imaging sequences consisted of the black blood double inversion recovery sequence (T1 weighted image and T2 weighted image) and breath-holding black blood three-inversion recovery sequence (FST1 weighted image and FST2 weighted image). All sequences were performed across different TG values, including 150, 160, 170, 180, and 190. The images’ quality was scored by two experienced radiologists using a four-point system based on the blinding principle. The paired variance analysis was made to compare the image quality of different TG groups. 【Results】 We observed consistent results in both normal and overweight groups. The optimal TG value was 160 for black blood triple inversion recovery sequence and 170 for black blood double inversion recovery sequence. 【Conclusion】 Adjusting the TG value can improve the signal uniformity of the myocardium, thereby improving the quality of images.

SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA