Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Añadir filtros








Asunto principal
Intervalo de año
1.
Korean Journal of Orthodontics ; : 171-181, 2007.
Artículo en Coreano | WPRIM | ID: wpr-650779

RESUMEN

OBJECTIVE: The present study was aimed at an analytical formulation of the micro-implant related torque as a function of implant size, i.e. the diameter and length, screw size, and the bony resistance at the implant to bone interface. METHODS: The resistance at the implant to cancellous bone interface (S(can)) was assumed to be in the range of 1.0-2.5 MPa. Micro-implant model of Absoanchor (Dentos Inc. Daegu, Korea) was used in the course of the analysis. RESULTS: The results showed that the torque was a strong function of diameter, length, and the screw height. As the diameter increased and as the screw size decreased, the torque index decreased. However the strength index was a different function of the implant and bone factors. The whole Absoanchor implant models were within the safe region when the resistance at the implant/cancellous bone (= S(can)) was 1.0 or less. CONCLUSION: For bone with S(can) of 1.5 MPa, the cervical diameter should be greater than 1.5 mm if micro-implant models of 12 mm long are to be placed. For S(can) of 2.0 MPa, micro-implant models of larger cervical diameter than 1.5 mm were found to be safe only if the endosseous length was less than 8 mm.


Asunto(s)
Torque
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA