Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Añadir filtros








Intervalo de año
1.
Journal of Laboratory Medicine and Quality Assurance ; : 247-250, 2003.
Artículo en Coreano | WPRIM | ID: wpr-119811

RESUMEN

BACKGROUND: Fecal occult blood tests (FOBT) have been recommended for gastro- intestinal bleeding and colon cancer screening. This study compared the effectiveness of two fecal blood screening kit, OC-Hemodia II and Ez step FOB for fecal occult blood. METHODS: The detection limit was evaluated by using OC-control and serially diluted samples. The comparison study between OC-hemodia II and Ez step FOB were evaluated in 143 cases. RESULTS: The concordance rate between OC-Hemodia II and Ez step FOB was 85.6% and 50% of non-concordance cases have history related to gatrointestinal bleeding. Ez step FOB was possible to detect 35 ng/mL in serially diluted OC-control. CONCLUSIONS: The result of Ez step FOB satisfactory to clinical application and showed good concordance rate compared to OC-Hemodia II.


Asunto(s)
Neoplasias del Colon , Hemorragia , Límite de Detección , Tamizaje Masivo , Sangre Oculta
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA