Chinese Journal of Radiation Oncology ; (6): 704-710, 2023.
Artículo en Chino | WPRIM | ID: wpr-993251

RESUMEN

Objective:To reconstruct the dose of nasopharyngeal carcinoma and verify the results of the whole-process radiotherapy plan based on log files and cone beam CT (CBCT).Methods:A total of 15 patients with nasopharyngeal carcinoma who received volumetric modulated arc therapy (VMAT) with Halcyon accelerator in the Cancer Center of Union Hospital, Tongji Medical College, Huazhong University of Science and Technology from February to September 2022 were retrospectively selected. Log files and CBCT for all fractionated radiotherapy were recorded. The errors of monitor unit (MU), gantry angle, and multi-leaf collimator (MLC) leaf position per control point were analyzed. The adaptive CT (aCT) were generated according to CBCT and planned CT (pCT) using a commercial software Velocity TM, and the similarities among aCT, pCT and CBCT were analyzed. The original plan was modified from the log files and imported into the treatment planning system to calculate the delivered dose on the corresponding fractionated aCT to reconstruct the fractionated dose. And all the reconstructed doses were mapped back to pCT to obtain the cumulative dose. Theγpass ratios with criteria of 2 mm/2% and 2 mm/3% and the dose differences between the planned dose and the cumulative dose in the planning target volume (PTV) and organs at risk (OAR) were compared. Results:The root mean square (RMS) and the 95th percentile of the errors of MU, gantry angle and MLC leaf position errors were within an acceptable range. The aCT generated by Velocity TM had the anatomical structure of CBCT and the resolution, contrast, noise characteristics of pCT, which could be directly used for dose calculation. Compared with the planned dose, the changes of V 70 Gy of nasopharyngeal primary tumor (PTV nx), V 68 Gy of cervical glands (PTV nd) and V 60 Gy of planning target volume (PTV1) were -0.88%±1.91%, -2.99%±2.99% and -0.63%±0.93%, respectively, and V 40 Gy of parotid gland was increased to 2.65%±2.63%. Cumulative dose showed different degrees of PTV dose decrease ( P<0.05) and parotid dose was increased ( P<0.05). The γ pass ratio (2 mm/3%) between the cumulative dose and planned dose was 97.3%±2.7% and >95.0% in 86.7% of patients. Conclusions:Based on the log files and CBCT, the whole-process dose reconstruction of nasopharyngeal carcinoma patients can be carried out. According to the results of dose reconstruction, the radiotherapy effect of the target area and OAR can be quantitatively evaluated. In the case of high dose coverage and conformity of the original plan, the reconstruction results show that the cumulative dose coverage of the target area is decreased, whereas that of the parotid gland is increased.

2.
Chinese Journal of Radiation Oncology ; (6): 712-716, 2021.
Artículo en Chino | WPRIM | ID: wpr-910455

RESUMEN

Objective:To understand the components, performance, acceptance test and quality assurance (QA) protocols of the new Varian Halcyon accelerator through the procedures of installation product acceptance (IPA).Methods:Per Varian IPA protocol, TG-142 for LINAC and TG-148 for tomotherapy QA protocols, the software license, safety interlock, mechanical accuracy, dosimetric performance and imager system were checked thoroughly. Some parameters were benchmarked to the conventional TrueBeam system.Results:The system has been fully licensed. Safety interlock was normal. Mechanical accuracy: The maximum deviation of beam stability per gantry rotation was 1.13%. The size of isocenter was>0.59mm. The offsets of MV imager, collimation rotation and absolute gantry rotation were 0.09mm, -0.21° and 0.11°, respectively. The maximum offsets of couch, virtual-to-isocenter were 0.15mm (vertical) and -0.04mm (vertical), respectively. Beam performance: The depth deviation of maximum dose was 0.1cm. The offset of percentage depth dose at 10cm was 0.5%. The maximum deviations of off-axis-intensity, symmetry, and repeatability were 0.9%, 0.94% and -0.44%, respectively. MV imager: The dark field mean pixel value, noise, corrected pixels, defective lines, sensitivity and linearity disparity of dose were 614, 4.4, 3626, 0, 19177, and 0.47%, respectively. All values were within the range of tolerance. Visual check of contrast resolution and small object detection was all satisfactory.Conclusions:Without Halcyon-specific TG report or guidelines, manufacture-provided IPA manual can be helpful with the installation of acceptance and QA protocols. IPA has been successfully performed for Halcyon at Peking University Cancer Hospital. The automated workflow improves the clinical efficiency by simplifying the operations.

SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA