China Pharmacist ; (12): 324-325,326, 2015.
Artículo en Chino | WPRIM | ID: wpr-600509

RESUMEN

Objective:To establish a method for the determination of chloride and sulfate in lithium carbonate crude drug by sup-pressed ion chromatography. Methods:The analysis column and the guard column was Thermo IonPac AS11(250 mm × 4 mm) and IonPac AG 11(50 mm ×4 mm), respectively. Potassium hydroxide solution(10 mmol·L-1) generated by an eluent generator(EG) was used as the eluent at a flow rate of 1. 0 ml·min-1 . And the temperature of column and conductivity cell was 30℃ and 35℃, re-spectively. The suppressor electric current was 36 mA, and the injection volume was 10 μl. Results:The linear range of chloride ion and sulfate ion was 1. 0-20μg·ml-1 with r of 0. 995 1 and 0. 997 3,respectively, and the recovery was 106. 6%(RSD=2. 6%,n=6) and 100. 3%(RSD=1. 9%,n=6), respectively. Conclusion:The method is accurate and reliable, and can be used in the determina-tion of chloride and sulfate in lithium carbonate crude drug.