Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 16 de 16
Filtrar
Añadir filtros








Intervalo de año
1.
Chinese Journal of Radiological Medicine and Protection ; (12): 145-149, 2020.
Artículo en Chino | WPRIM | ID: wpr-868416

RESUMEN

Objective To study a method for verifying the doses to PTV and OAR as well as the 2D dose distribution arising from IMRT through using radiochromic films and TLDs.Methods Totally 7 medical electronic linear accelerators from Varian,Siemens and Elekta were selected.The polystyrene phantom provided by IAEA was conducted with CT scan.After irradiation with 6 MV X-rays,the TLDs and films were returned to the secondary standard dosimetry laboratory of China CDC for measurement and estimation.Results According to the IAEA requirements,the relative deviations between TLD-measured and TPS-planned values for PTV and OAR doses were both within ±7.0%.For PTV,the measured relative deviations for 5 accelerators were in the range of-4.0% to 3.4%,consistent with the IAEA requirements,whereas the values for the other 2 accelerators were in the range of-7.0% to 10.6%,not consistent with the requirements.For OAR,the values for 4 accelerators were in the range of-5.6% to 3.3%,consistent with the IAEA requirements,whereas the values for the other 3 accelerators were in the range of-20.8% to 11.5%,not meeting the requirements.As required by the IAEA,the 2D dose distribution 3 mm/3% pass rate should be higher than 90%.The measured values for 5 accelerators were in the range of 91.8% to 98.5%,consistent with the requirements,whereas the values measured for the other 2 were 45.0% and 77.0% respectively,not meeting the requirements.Conclusions It is feasible for using TLDs and radiochromic films to verify the doses to PTV and OAR and the 2D dose distribution in IMRT.This method should be applied to not only quality verification but also hospital internal audit to the extent possible.

2.
Chinese Journal of Radiological Medicine and Protection ; (12): 140-144, 2020.
Artículo en Chino | WPRIM | ID: wpr-868415

RESUMEN

Objective To use TLDs and radiochromic films to verify the prescribed doses to both planned target volume (PTV) and organ at risk (OAR) and the 2D dose distribution in IMRT.Methods Eight accelerators of different models were selected in Henan province.The polystyrene phantom provided by IAEA was scanned using CT scanners and then the scanned images were transmitted to treatment planning system (TPS) for prescribing respectively the doses to PTV and OAR.IMRT was performed with phantom exposed to a 6 MV X-rays.The irradiated TLDs and films were delivered for measurement and estimation at Secondary Standard Dosimetry Laboratory at National Institute for Radiological Protection,Chinese Center for Disease Control and Prevention.Results According to IAEA requirements,the relative deviations of the TLD-measured and TPS-planned values were within ±7.0% for the prescribed doses to PTV and OAR.The measured results for PTV have shown that the relative deviation of TLD-measured and TPS-planned values were within-0.3% to 6.9% for 8 accelerators,all consistent with the IAEA requirements.For OAR,the relative deviations of TLD-measured and TPS-planned were within-7.0% to 0.3% for 6 accelerators,consistent with the requirements,whereas those for other 2 accelerators were within-10.8% to-8.4%,not up to the requirements.IAEA required that,for 2D dose distribution,the pass rate of 3 mm/3% be ≥ 90%.The measured values for 7 accelerators were from 90.2% to 99.9%,consistent with the requirements,whereas that for another one was 70.0%,not meeting the requirement.Conclusions The method to verify,using radiochromic film and TLD,the prescribed doses to PTV and OAR and the pass rate of 2D dose distribution is simple and reliable.It is an important step to implement quality control for IMRT and can provide effective support for medical or third-party service institution to verify clinically prescribed dose.

3.
Chinese Journal of Radiological Medicine and Protection ; (12): 134-139, 2020.
Artículo en Chino | WPRIM | ID: wpr-868414

RESUMEN

Objective To study a method to measure the doses to planned target volume (PTV) and organ at risk (OAR) and 2D dose distribution in IMRT by using TLD and radiochromic film for a verification purpose.Methods Totally 7 different types of medical linear accelerators were selected from seven hospitals in Hubei province.A polystyrene phantom provided by IAEA was CT scanned and then the scanned images were returned to the Treatment Planning System (TPS) for determining the prescribed doses to PTS and OAR and the corresponding MU.After the phantom was irradiated with 6 MV X-ray,the TLDs and films were transmitted to the secondary standard dosimetry laboratory of China CDC for measurement and estimation.Results The IAEA required the relative deviations between TLD-measured and TPS-planned doses to OAR and PTV be within ±7.0%.For PTV,the measured-to-planned deviation values for 7 accelerator were within-5.4% to 6.5%,all consistent with the IAEA requirements.For OAR,the values for 5 accelerators were within-2.2% to 6.7%,not consistent the requirements,whereas the values for the other 2 were-8.6% and 8.2% respectively,beyond the required values.The IAEA required that the 2D dose distribution 3 mm/3% pass rate be higher than 90%.The measured values for 7 accelerators were in the range of 90.3%-98.9%,all consistent with the requirements.Conclusions It is feasible scientifically and easy to operate in practice for using TLD and film to carry out dose verification in IMRT.It would be advisable to apply this method to quality verification in IMRT in medical institutions to the extent possible.

4.
Chinese Journal of Radiological Medicine and Protection ; (12): 129-133, 2020.
Artículo en Chino | WPRIM | ID: wpr-868413

RESUMEN

Objective To validate the method for measuring the TPV and OAR doses and 2D dose distribution in IMRT through using TLD and radiochromic film.Methods Eight medical linear accelerators (Valian,Elekta,Siemens) were selected.The polystyrene phantom provided by IAEA was CT scanned and the image obtained was transferred to TPS for formulation of treatment plan,prescription of PTV and OAR doses and calculation of corresponding monitoring unit (MU),IMRT was performed on the phantom using 6 MV X-ray.Irradiated TLDs and films were measured and evaluated at the Secondary Standard Dosimetry Laboratory at the Radiation Safety Institute of Chinese Center for Disease Control and Prevention.Results According to IAEA requirement,the relative deviations between TLD-measured and TPS-planned doses were within ±7.0% for the prescribed PTV and OAR doses.As measured result,the PTV values for 8 accelerators were in the range of 0.6% to 5.9%,consistent with the IAEA requirements,whereas the OAT values for 8 accelerators were within-0.6% to 7.0%,consistent the requirements.As IAEA required,the 2D dose distribution passing rate of 3 mm/3% should be higher than 90%.The filmmeasured and TPS-planned values for 8 accelerators were within 90.2% to 100.0%,consistent with the requirements.Conclusions TLD and radiochromic film are feasible in validating the PTV and OAR doses and the 2D dose distribution pass rate in IMRT.This method can be widely used in quality audit and internal verification in IMRT in medical institutiions on a large scale.

5.
Chinese Journal of Radiological Medicine and Protection ; (12): 122-128, 2020.
Artículo en Chino | WPRIM | ID: wpr-868412

RESUMEN

Objective To develop the methodology for using TLD and radiochromic film to measure the planned target volume (PTV) and organ at risk (OAR) doses and 2D dose distribution in IMRT,in order to provide technical guidance on the dose quality audit in IMRT at home.Methods China has participated in the research project launched by the international multi-radiotherapy centre (IMRC).IMRT polystyrene phantom provided by IAEA was scanned by CT scanner and then the scanned images were transmitted to TPS to outline prescribed dose to PTV and to OAR.The former was limited to 400 cGy while the latter limited to 200 cGy.IMRT was implemented with the phantom irradiated using 6 MV X-ray.The irradiated TLDs and films were sent to IAEA dosimerty laboratory for measurement and calculation.Jiangsu,Sichuan,Hubei and Henan provinces were selected to engage with this study for their variety of accelerators and highly skilled physicists.The procedures used were the same as in the IMRC and the irradiated TLDs and films were required to send to external audit group for measurement and calculation.Results According to IAEA requirement,the relative deviations of the TLD-measured and TPS planned doses are within ±7.0% for PTV and OAR.The China's research results at the IMRC have shown that the relative deviation of TLD-measured and TPS-planned values for the upper and lower PTV were-0.2% and 0.8%,respectively,consistent with the IAEA requirement,and the values for upper and lower OAR were -0.6% and-1.0%,respectively,consistent with the requirement.As the results have shown in four provinces,the relative deviations of the TLD-measured and TPS-planned were within 0 to 10.6% for upper and lower PTV and-0.6% to 20.9% for upper and lower OAR.According to IAEA requirement,the passing rate should be greater than 90% for 3 mm /3% for 2D dose distribution.China's result at the IMRC is 100%,being excellent.The four provinces' results have shown that 2D dose distribution pass rate of 3 mm/3% was in the range of 45.0%-100.0%.Conclusions The uses of TLD in quality audit for PTV and OAR doses and the radiochromic film in 2D dose distribution pass rate in IMRT are characterized by scientific feasibility,strong operability,easy-to-mail and data realibility.They are can be applied to quality assurance and audit in medical institutions in the country to on a large scale.

6.
Chinese Journal of Radiological Medicine and Protection ; (12): 145-149, 2020.
Artículo en Chino | WPRIM | ID: wpr-799421

RESUMEN

Objective@#To study a method for verifying the doses to PTV and OAR as well as the 2D dose distribution arising from IMRT through using radiochromic films and TLDs.@*Methods@#Totally 7 medical electronic linear accelerators from Varian, Siemens and Elekta were selected. The polystyrene phantom provided by IAEA was conducted with CT scan. After irradiation with 6 MV X-rays, the TLDs and films were returned to the secondary standard dosimetry laboratory of China CDC for measurement and estimation.@*Results@#According to the IAEA requirements, the relative deviations between TLD-measured and TPS-planned values for PTV and OAR doses were both within ±7.0%. For PTV, the measured relative deviations for 5 accelerators were in the range of -4.0% to 3.4%, consistent with the IAEA requirements, whereas the values for the other 2 accelerators were in the range of -7.0% to 10.6%, not consistent with the requirements. For OAR, the values for 4 accelerators were in the range of -5.6% to 3.3%, consistent with the IAEA requirements, whereas the values for the other 3 accelerators were in the range of -20.8% to 11.5%, not meeting the requirements. As required by the IAEA, the 2D dose distribution 3 mm/3% pass rate should be higher than 90%. The measured values for 5 accelerators were in the range of 91.8% to 98.5%, consistent with the requirements, whereas the values measured for the other 2 were 45.0% and 77.0% respectively, not meeting the requirements.@*Conclusions@#It is feasible for using TLDs and radiochromic films to verify the doses to PTV and OAR and the 2D dose distribution in IMRT. This method should be applied to not only quality verification but also hospital internal audit to the extent possible.

7.
Chinese Journal of Radiological Medicine and Protection ; (12): 140-144, 2020.
Artículo en Chino | WPRIM | ID: wpr-799420

RESUMEN

Objective@#To use TLDs and radiochromic films to verify the prescribed doses to both planned target volume (PTV) and organ at risk (OAR) and the 2D dose distribution in IMRT.@*Methods@#Eight accelerators of different models were selected in Henan province. The polystyrene phantom provided by IAEA was scanned using CT scanners and then the scanned images were transmitted to treatment planning system (TPS) for prescribing respectively the doses to PTV and OAR. IMRT was performed with phantom exposed to a 6 MV X-rays. The irradiated TLDs and films were delivered for measurement and estimation at Secondary Standard Dosimetry Laboratory at National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention.@*Results@#According to IAEA requirements, the relative deviations of the TLD-measured and TPS-planned values were within ±7.0% for the prescribed doses to PTV and OAR. The measured results for PTV have shown that the relative deviation of TLD-measured and TPS-planned values were within -0.3% to 6.9% for 8 accelerators, all consistent with the IAEA requirements. For OAR, the relative deviations of TLD-measured and TPS-planned were within -7.0% to 0.3% for 6 accelerators, consistent with the requirements, whereas those for other 2 accelerators were within -10.8% to -8.4%, not up to the requirements. IAEA required that, for 2D dose distribution, the pass rate of 3 mm/3% be ≥90%. The measured values for 7 accelerators were from 90.2% to 99.9%, consistent with the requirements, whereas that for another one was 70.0%, not meeting the requirement.@*Conclusions@#The method to verify, using radiochromic film and TLD, the prescribed doses to PTV and OAR and the pass rate of 2D dose distribution is simple and reliable. It is an important step to implement quality control for IMRT and can provide effective support for medical or third-party service institution to verify clinically prescribed dose.

8.
Chinese Journal of Radiological Medicine and Protection ; (12): 134-139, 2020.
Artículo en Chino | WPRIM | ID: wpr-799419

RESUMEN

Objective@#To study a method to measure the doses to planned target volume (PTV) and organ at risk (OAR) and 2D dose distribution in IMRT by using TLD and radiochromic film for a verification purpose.@*Methods@#Totally 7 different types of medical linear accelerators were selected from seven hospitals in Hubei province. A polystyrene phantom provided by IAEA was CT scanned and then the scanned images were returned to the Treatment Planning System (TPS) for determining the prescribed doses to PTS and OAR and the corresponding MU. After the phantom was irradiated with 6 MV X-ray, the TLDs and films were transmitted to the secondary standard dosimetry laboratory of China CDC for measurement and estimation.@*Results@#The IAEA required the relative deviations between TLD-measured and TPS-planned doses to OAR and PTV be within ±7.0%. For PTV, the measured-to-planned deviation values for 7 accelerator were within -5.4% to 6.5%, all consistent with the IAEA requirements. For OAR, the values for 5 accelerators were within -2.2% to 6.7%, not consistent the requirements, whereas the values for the other 2 were -8.6% and 8.2% respectively, beyond the required values. The IAEA required that the 2D dose distribution 3 mm/3% pass rate be higher than 90%. The measured values for 7 accelerators were in the range of 90.3%-98.9%, all consistent with the requirements.@*Conclusions@#It is feasible scientifically and easy to operate in practice for using TLD and film to carry out dose verification in IMRT. It would be advisable to apply this method to quality verification in IMRT in medical institutions to the extent possible.

9.
Chinese Journal of Radiological Medicine and Protection ; (12): 129-133, 2020.
Artículo en Chino | WPRIM | ID: wpr-799418

RESUMEN

Objective@#To validate the method for measuring the TPV and OAR doses and 2D dose distribution in IMRT through using TLD and radiochromic film.@*Methods@#Eight medical linear accelerators (Valian, Elekta, Siemens) were selected. The polystyrene phantom provided by IAEA was CT scanned and the image obtained was transferred to TPS for formulation of treatment plan, prescription of PTV and OAR doses and calculation of corresponding monitoring unit (MU), IMRT was performed on the phantom using 6 MV X-ray. Irradiated TLDs and films were measured and evaluated at the Secondary Standard Dosimetry Laboratory at the Radiation Safety Institute of Chinese Center for Disease Control and Prevention.@*Results@#According to IAEA requirement, the relative deviations between TLD-measured and TPS-planned doses were within ±7.0% for the prescribed PTV and OAR doses. As measured result, the PTV values for 8 accelerators were in the range of 0.6% to 5.9%, consistent with the IAEA requirements, whereas the OAT values for 8 accelerators were within -0.6% to 7.0%, consistent the requirements. As IAEA required, the 2D dose distribution passing rate of 3 mm/3% should be higher than 90%. The film-measured and TPS-planned values for 8 accelerators were within 90.2% to 100.0%, consistent with the requirements.@*Conclusions@#TLD and radiochromic film are feasible in validating the PTV and OAR doses and the 2D dose distribution pass rate in IMRT. This method can be widely used in quality audit and internal verification in IMRT in medical institutiions on a large scale.

10.
Chinese Journal of Radiological Medicine and Protection ; (12): 122-128, 2020.
Artículo en Chino | WPRIM | ID: wpr-799417

RESUMEN

Objective@#To develop the methodology for using TLD and radiochromic film to measure the planned target volume (PTV) and organ at risk (OAR) doses and 2D dose distribution in IMRT, in order to provide technical guidance on the dose quality audit in IMRT at home.@*Methods@#China has participated in the research project launched by the international multi-radiotherapy centre (IMRC). IMRT polystyrene phantom provided by IAEA was scanned by CT scanner and then the scanned images were transmitted to TPS to outline prescribed dose to PTV and to OAR. The former was limited to 400 cGy while the latter limited to 200 cGy. IMRT was implemented with the phantom irradiated using 6 MV X-ray. The irradiated TLDs and films were sent to IAEA dosimerty laboratory for measurement and calculation. Jiangsu, Sichuan, Hubei and Henan provinces were selected to engage with this study for their variety of accelerators and highly skilled physicists. The procedures used were the same as in the IMRC and the irradiated TLDs and films were required to send to external audit group for measurement and calculation.@*Results@#According to IAEA requirement, the relative deviations of the TLD-measured and TPS planned doses are within ±7.0% for PTV and OAR. The China′s research results at the IMRC have shown that the relative deviation of TLD-measured and TPS-planned values for the upper and lower PTV were -0.2% and 0.8%, respectively, consistent with the IAEA requirement, and the values for upper and lower OAR were -0.6% and -1.0%, respectively, consistent with the requirement. As the results have shown in four provinces, the relative deviations of the TLD-measured and TPS-planned were within 0 to 10.6% for upper and lower PTV and -0.6% to 20.9% for upper and lower OAR. According to IAEA requirement, the passing rate should be greater than 90% for 3 mm /3% for 2D dose distribution. China′s result at the IMRC is 100%, being excellent. The four provinces′ results have shown that 2D dose distribution pass rate of 3 mm/3% was in the range of 45.0%-100.0%.@*Conclusions@#The uses of TLD in quality audit for PTV and OAR doses and the radiochromic film in 2D dose distribution pass rate in IMRT are characterized by scientific feasibility, strong operability, easy-to-mail and data realibility. They are can be applied to quality assurance and audit in medical institutions in the country to on a large scale.

11.
Chinese Journal of Radiological Medicine and Protection ; (12): 137-140, 2019.
Artículo en Chino | WPRIM | ID: wpr-734329

RESUMEN

Objective To measure the positioning accuracy of multi-leaf collimators (MLC) leaves by using radiochromic films,with aim to providing data in support of IAEA's methodology validation.Methods The present study focused on 8 accelerators (Varian,Elekta and Siemens machines) at 6 hospitals in Henan province with highly skilled physicists.The 25 cm × 25 cm radiochromic films were put on 30 cm × 30 cm homogeneous solid phantoms and covered with a 2.0-cm-thick homogeneous solid phantom slabs.The CT-scanned images were transmitted to TPS for plan formulation.With 6 MV X-ray,MLC created 5 strip 3 cm × 0.6 cm picket fence field,each with 3.0 cm strip separation.The SSD was 100 cm at the maximum dose point,with a 250 MU to each strip.The irradiated radiochromic films were returned to IAEA for analysis within one week and compared with the values given by IAEA.Results The difference of film-measured and TPS-planned positions of MLC leaves for each strip picket fence should be within ± 0.5 mm as required by IAEA.For 7 of 8 accelerators selected,the differences of accurately measured MLC leaf positions were all within ±0.5 mm,which were in line with the IAEA requirements,with only other one being beyond-0.5 mm,not consistent with the IAEA requirements.The differences of film-measured actual widths between each pair and all pairs of leaves were within ±0.75 mm as required by IAEA for 7 accelerators,whereas the other one was outside ± 0.75 mm,not consistent with the IAEA requirements.The standard deviation of film-measured actual width of MCL leaf between each pair and all pairs for 7 accelerators were ≤ 0.3 mm as required by IAEA,whereas the other one was 0.5 mm,not consistent with IAEA requirements.Conclusions The MLC positioning accuracy of a few of medical accelerators in Henan is not qualified.It is of great significance to carry out regular quality examination as well as third-party verification to ensure precise delivery of radiotherapy.

12.
Chinese Journal of Radiological Medicine and Protection ; (12): 132-136, 2019.
Artículo en Chino | WPRIM | ID: wpr-734328

RESUMEN

Objective To verify the leaf positioning accuracy of multi-leaf collimator (MLC) in intensity modulated radiation therapy (IMRT) by using radiochromic films to provide references for IMRT quality control.Methods Medical linear accelerators of different designs,owned by 7 first-class hospitals at grade-3 in Hubei,were selected for this verification study.Five strip picket fence pattern was created by treatment planning system (TPS).Each strip was 0.6 cm wide with a 3 cm strip separation.MLC leaf positions,positioning bias and opening widths were then measured with EBT2 radiochromic films.Results According to IAEA standards,the film-measured and TPS planned MLC leaf position difference was ±0.5 mm.The difference of MCL strip position of No.5 and No.7 accelerator was 0.7 and-1.0 mm,respectively,not in line with the IAEA standards.The difference of film-measured MLC leaf position between each pairs and all pairs of leaves of 7 accelerators were all within ± 0.5 mm,in line with the IAEA requirements.As required by IAEA,the difference of opening width of MLC leave of each pair relative to the averaged widths of all pairs should be within ±0.75 mm.The filem-measured values from 7 accelerators ranged from-0.6 to 0.5 mm,all in line with the IAEA standards.Standard deviation of opening width of all leaves were required to be within 0.3 mm.The measured values from 7 accelerators ranged from 0.1 to 0.2 mm,in line with the IAEA standard.Conclusions The use of EBT2 radiochromic films to verify MLC leaf positioning accuracy is an important means of quality control owing to its simplicity and high measurement accuracy.It is recommended for verification use at a large scale.

13.
Rev. mex. ing. bioméd ; 38(1): 363-371, ene.-abr. 2017. graf
Artículo en Español | LILACS-Express | LILACS | ID: biblio-902355

RESUMEN

Resumen: Objetivo: Analizar los espectros de absorción neta de las películas radiocrómicas EBT2 y EBT3 para describir su influencia en el comportamiento de las curvas de dosis-respuesta. Metodología: Las películas se irradiaron en un acelerador lineal de 6 MV. La obtención de los espectros de absorción neta se realizó con espectrofotómetro UV/VIS. Las curvas de dosis-respuesta se obtuvieron con un escáner, un láser He-Ne y un espectrofotómetro. Resultados: El espectro de absorción de las EBT2 muestra tres bandas de absorción centradas que conservan la posición y aumentan su intensidad en función de la dosis, sin embargo, este comportamiento no se observa en las películas EBT3. La curva dosis-respuesta muestra la máxima sensibilidad utilizando el espectrofotómetro, pero no muestra un comportamiento definido. Implicaciones: Generación de nuevos conocimientos para la creación de nuevos sistemas ópticos capaces de amplificar la sensibilidad de la respuesta de las películas. Originalidad: Mostrar la correlación entre los espectros de absorción neta y su influencia en las curvas dosis-respuesta en tres diferentes sistemas ópticos. Conclusiones: El comportamiento de los espectros de absorción aunado al comportamiento de las curvas dosis-respuesta nos ayuda a descartar el uso de sistemas ópticos que no garanticen un uso clínico confiable.


Abstract: Objective: To analyze the net absorption spectra of EBT2 and EBT3 radiochromic films to describe their influence on the behavior of dose-response curves. Methodology: The films were irradiated in a linear accelerator of 6 MV. The net absorption spectra were obtained with a UV / VIS spectrophotometer. Dose-response curves were obtained with a scanner, a He-Ne laser and a spectrophotometer. Results: The absorption spectrum of the EBT2 shows three focused absorption bands that retain position and increase their intensity as a function of dose, however, this behavior is not observed in EBT3 films. The dose-response curve shows maximum sensitivity using the spectrophotometer, but does not show a defined behavior. Implications: Generation of new knowledge for the creation of new optical systems capable of amplifying the responsiveness of the films. Originality: Show the correlation between net absorption spectra and their influence on dose-response curves in three different optical systems. Conclusions: The behavior of absorption spectra combined with the behavior of the dose-response curves helps to discard the use of optical systems that do not guarantee a reliable clinical use.

14.
Braz. j. med. biol. res ; 50(3): e5848, 2017. tab, graf
Artículo en Inglés | LILACS | ID: biblio-839264

RESUMEN

This study presents the characterization of an X-ray irradiator through dosimetric tests, which confirms the actual dose rate that small animals and cells will be exposed to during radiobiological experiments. We evaluated the linearity, consistency, repeatability, and dose distribution in the positions in which the animals or cells are placed during irradiation. In addition, we evaluated the performance of the X-ray tube (voltage and tube operating current), the radiometric survey (leakage radiation) and safety devices. The irradiator default setting was established as 160 kV and 25 mA. Tests showed that the dose rate was linear overtime (R2=1) and remained stable for long (constant) and short (repeatability) intervals between readings. The mean dose rate inside the animal cages was 1.27±0.06 Gy/min with a uniform beam of 95.40% (above the minimum threshold guaranteed by the manufacturer). The mean dose rate inside the cell plates was 0.92±0.19 Gy/min. The dose rate dependence with tube voltage and current presented a quadratic and linear relationship, respectively. There was no observed mechanical failure during evaluation of the irradiator safety devices and the radiometric survey obtained a maximum ambient equivalent dose rate of 0.26 mSv/h, which exempts it from the radiological protection requirements of the International Atomic Energy Agency. The irradiator characterization enables us to perform radiobiological experiments, and assists or even replaces traditional therapy equipment (e.g., linear accelerators) for cells and small animal irradiation, especially in early research stages.


Asunto(s)
Animales , Dosis de Radiación , Radiometría/instrumentación , Calibración , Diseño de Equipo , Aceleradores de Partículas , Radiometría/métodos , Rayos X
15.
Chinese Journal of Radiological Medicine and Protection ; (12): 145-148, 2014.
Artículo en Chino | WPRIM | ID: wpr-444860

RESUMEN

Objective To study the dosimetric characteristics of two different radiochromic films used to estimate peak skin dose(PSD) of patients.The characteristics of these two films were investigated and compared after exposure to ionizing radiation in the diagnostic energy range,including post-exposure gray growth,sensitivity,energy dependence and dose response.Methods GafChromic XR-RV3 film and KODAK EDR2 film were exposed to air kerma 800 mGy free-in-air using five X-ray beam qualities (60,80,100,120 and 140 kVp) in a SSDL.The measurement for each energy was normalized to 80 kV to analyze energy dependence of films.The films were calibrated to different dose level (0.025-10 Gy) onphantom by 80 kV X-rays.The response curve were plotted to analyze sensitivity and dose response.The films were scanned with Epson V750 commercial flatbed scanner.Color channel analysis was performed.Results The post-exposure gray growth of XR-RV3 film was found to be fairly stable.The change were 2%,4%,6% at 24 h,72 h and 6 weeks after exposure respectively.EDR2 film was found to be more sensitivity than XR-RV3 film in low dose.The energy response of the XR-RV3 film and EDR2 film were within 9% and 23% over the clinical diagnostic x-ray energies,respectively.In the dose range of 0.025-10 Gy,for the XR-RV3 film,the red channel with the dose response curve was most obvious.For EDR2 film,the pixel value of three color channels was coincident.The EDR2 film appeared to be saturated when receiving doses greater than 500 mGy.Conclusions The XR-RV3 film is superior to EDR2 film in gray growth,energy dependence,dose-response and other aspects.This film is very suitable for measuring and analyzing PSD of patients in interventional radiology procedures.

16.
Biomedical Imaging and Intervention Journal ; : 1-12, 2009.
Artículo en Inglés | WPRIM | ID: wpr-625887

RESUMEN

The authors have studied the feasibility of using three new high-sensitivity radiochromic devices in measuring the doses to peripheral points outside the primary megavoltage photon beams. The three devices were GAFCHROMIC® EBT film, prototype Low Dose (LD) Film, and prototype LD Card. The authors performed point dosimetry using these three devices in water-equivalent solid phantoms at x = 3,5,8,10, and 15 cm from the edge of 6 MV and 15 MV photon beams of 10x10 cm2, and at depths of 0, 0.5 cm, and depth of maximum dose. A full sheet of EBT film was exposed with 5000 MU. The prototype LD film pieces were 1.5x2 cm2 in size. Some LD films were provided in the form of a card in 1.8x5 cm2 holding an active film in 1.8x2 cm2. These are referred to as “LD dosimeter cards”. The small LD films and cards were exposed with 500 MU. For each scanned film, a 6 mm circular area centered at the measurement point was sampled and the mean pixel value was obtained. The calibration curves were established from the calibration data for each combination of film/cards and densitometer/scanner. The doses at the peripheral points determined from the films were compared with those obtained using ion chamber at respective locations in a water phantom and general agreements were found. It is feasible to accurately measure peripheral doses of megavoltage photon beams using the new high-sensitivity radiochromic devices. This near real-time and inexpensive method can be applied in a clinical setting for dose measurements to critical organs and sensitive patient implant devices.

SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA