Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 3 de 3
Filtrar
Añadir filtros








Intervalo de año
1.
Braz. j. pharm. sci ; 45(4): 701-708, Oct.-Dec. 2009. ilus, tab
Artículo en Inglés | LILACS | ID: lil-543666

RESUMEN

Efforts to diminish the transmission of infections include programs in which disinfectants play a crucial role. Hospital surfaces and medical devices are potential sources of cross contamination, and each instrument, surface or area in a health care unit can be responsible for spread of infection. The decimal reduction time was used to study and compare the behavior of selected strains of microorganisms. The highest D-values for various bacteria were obtained for the following solutions: (i) 0.1 percent sodium dichloroisocyanurate (pH 7.0) - E. coli and A. calcoaceticus (D = 5.9 min); (ii) sodium hypochlorite (pH 7.0) at 0.025 percent for B. stearothermophilus (D = 24 min), E. coli and E. cloacae (D = 7.5 min); at 0.05 percent for B. stearothermophilus (D = 9.4 min) and E. coli (D = 6.1 min). The suspension studies were an indication of the disinfectant efficacy on a surface. The data in this study reflect the formulations used and may vary from product to product. The expected effectiveness from the studied formulations shows that the tested agents can be recommended for surface disinfection as stated in present guidelines and emphasize the importance and need to develop routine and novel programs to evaluate product utility.


Esforços para diminuir o risco de transmissões de infecções incluem programas nos quais os desinfetantes desempenham papel crucial. As superfícies de materiais médico-hospitalares, se não estiverem diretamente ligados à transmissão de doenças, podem contribuir, potencialmente, para uma contaminação cruzada secundária. Cada instrumento ou superfície do estabelecimento do ambiente de saúde que entra em contato com um paciente é um disseminador potencial de infecção. Para estudar e comparar o comportamento dos microrganismos selecionados foram realizados ensaios de determinação do tempo de redução decimal. Os maiores valores D determinados, foram: (i) 0,1 por cento dicloroisocianurato de sódio (NaDCC) (pH 7.0) - E. coli e A. calcoaceticus (D = 5,9 min); (ii) hipoclorito de sódio (pH 7,0) à 0,025 por cento para B. stearothermophilus (D = 24 min), E. coli e E. cloacae (D = 7,5 min); e à 0,05 por cento para B. stearothermophilus (D = 9,4 min) e E. coli (D = 6,1 min). Este estudo estabelece que as suspensões estudadas são indicação da eficácia de desinfecção recomendada pela legislação, mas os resultados podem variar de produto para produto. Para desinfecção de mãos clorexidina pode ser utilizada, pois apresentou valores D baixos. Para desinfecção de nível intermediário de equipamentos e instrumentos recomenda-se a utilização de NaDCC, devido à estabilidade e baixo efeito corrosivo para equipamentos e materiais. Glutaraldeído, apesar de muito aceito para processos esterilizantes, tem eficácia comparável a soluções de formaldeído.


Asunto(s)
Escala Decimal , Diagnóstico , Desinfección/métodos , Esterilización/métodos , Fenómenos Químicos/prevención & control , Factores de Tiempo
2.
Artículo en Inglés | LILACS-Express | LILACS, VETINDEX | ID: biblio-1469426

RESUMEN

Prior to selecting disinfectant for low, intermediate and high (sterilizing) levels, the decimal reduction time, D-value, for the most common or persistent bacteria identified on a medical device or at a health care facility should be determined. The D-value was determined by inoculating 100 mL of disinfecting solution with 1 mL of a bacterial suspension. At regular intervals, 1 mL aliquots of this mixture were transferred to 8 mL of growth media containing a neutralizing agent, and incubated at optimal conditions for the microorganism. B. stearothermophilus and E. coli were the most resistant bacteria for the disinfecting and sterilizing procedures.


Para selecionar o agente sanitizante de acordo com o nível (baixo, intermediário, alto ou esterilizante) é necessário determinar o tempo de redução decimal (valor D) para os microrganismos comumente identificados em equipamentos médico-hospitalares. O valor D é determinado inoculando-se 1 mL da suspensão de microrganismo em 100 mL da solução desinfetante. Em intervalos constantes, alíquotas de 1 mL da mistura devem ser transferidas para 8 mL de meio de cultura contendo agente neutralizante. B. stearothermophilus e E. coli se mostraram os microrganismos mais resistentes para soluções esterilizantes e desinfetantes.

3.
Artículo en Inglés | LILACS-Express | LILACS, VETINDEX | ID: biblio-1469473

RESUMEN

Prior to selecting disinfectant for low, intermediate and high (sterilizing) levels, the decimal reduction time, D-value, for the most common or persistent bacteria identified on a medical device or at a health care facility should be determined. The D-value was determined by inoculating 100 mL of disinfecting solution with 1 mL of a bacterial suspension. At regular intervals, 1 mL aliquots of this mixture were transferred to 8 mL of growth media containing a neutralizing agent, and incubated at optimal conditions for the microorganism. B. stearothermophilus and E. coli were the most resistant bacteria for the disinfecting and sterilizing procedures.


Para selecionar o agente sanitizante de acordo com o nível (baixo, intermediário, alto ou esterilizante) é necessário determinar o tempo de redução decimal (valor D) para os microrganismos comumente identificados em equipamentos médico-hospitalares. O valor D é determinado inoculando-se 1 mL da suspensão de microrganismo em 100 mL da solução desinfetante. Em intervalos constantes, alíquotas de 1 mL da mistura devem ser transferidas para 8 mL de meio de cultura contendo agente neutralizante. B. stearothermophilus e E. coli se mostraram os microrganismos mais resistentes para soluções esterilizantes e desinfetantes.

SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA