1.
Chinese Journal of Radiological Medicine and Protection ; (12): 285-288, 2012.
Artículo en Chino | WPRIM | ID: wpr-427091

RESUMEN

Objective To verify the dose delivery accuracy of volumetric-modulated arc therapy plan by log-file analysis of linear accelerator that can be created when a dynamic delivery occurs.Methods Accelerator log file in binary format recorded the accelerator execution plan for each control point corresponding to the gantry angle,multi-leaf collimator leave position,cumulative machine monitor units ( MU).These information were read from the accelerator log file with Matlab7.1,then the original control points in the plan file replaced the corresponding information for the log,which generated a new plan.New plan was exported into the planning system to reculculate the dose.The volume dose histogram (DVH) and dose distribution was contrasted to determine the accuracy of the accelerator plan of implementation between two plans.Results Compared with the original plan,antry angle difference over ± 1° accounted for about 35% of the entire arc of control points in 4 of 12 arcs and the percentage of the leave error of ±0.5 mm was about 95%.MU error of a single control point was larger,but the cumulative MU for each are was small which was located between-0.09% to 0.11% in the selected 12 arcs.Between the targets,the maximum dose,minimum dose,the mean dose differences were from-0.07% to 0.42%,-0.38% to 0.40%,0.03% to 0.08%,respectively.The maximum dose and mean dose differences of organs at risks were located from-1.16% to 2.51%,-1.21% to 3.12%,respectively.Conclusions Accelerator log-file analysis to verify the VMAT plan nan be supplyed to the experimental method supplement.

2.
Chinese Journal of Radiological Medicine and Protection ; (12): 317-321, 2011.
Artículo en Chino | WPRIM | ID: wpr-416582

RESUMEN

Objective To compare the dosimetric difference between volumetric are modulation with RapidArc and fixed field dynamic IMRT for breast cancer radiotherapy after breast-conserving surgery.Methods Twenty patients with early left-sided breast cancer received radiotherapy after breast-conserving surgery.After target definition,treatment planning was performed by RapidAre and two fixed fields dynamic IMRT respectively on the same CT scan.The target dose distribution,homogeneity of the breast,and the irradiation dose and volume for the lungs,heart,and eontralateral breast were read in the dosevolume histogram (DVH) and compared between RapidAre and IMRT.The treatment delivery time and monitor units were also compared.Results In comparison with the IMRT planning,the homogeneity of clinical target volume (CTV) ,the volume proportion of 95% prescribed dose (V95%) was significantly higher by 0.65% in RapidAre (t =5.16,P = 0.001) ,and the V105% and V110% were lower by 10.96% and 1.48 % respectively,however,without statistical significance (t =-2.05 ,P =0.055 and t =-1.33 ,P =0.197).The conformal index of planning target volume (PTV) by the Rap~dAre planning was (0.88±0.02),significantly higher than that by the IMRT planning [(0.74±0.03),t = 18.54,P < 0.001].The homogeneity index (HI) of PTV by the RapidArc planning was 1.11±0.01,significantly lower than that by the IMRT planning (1.12±0.02,t =-2.44,P =0.02).There were no significant differences in the maximum dose (Dmax) and V20 for the ipsilateral lung between the RapidArc and IMRT planning,but the values of V10,V5 ,Dmin and Dmean by RapidArc planning were all significantly higher than those by the IMRT planning (all P < 0.01).The values of max dose and V30 for the heart were similar by both techniques,but the values of V10 and V5 by the RapidArc planning were significantly higher (by 18% and 50% ,respectively).The V5 of the contralateral breast and lung by the RapidArc planning were increased by 9.33% and 3.04% respectively compared to the IMRT planning.The mean MU of the RapidArc was 608 MU,significantly higher than that by the IMRT planning (437 MU,t = 10.86,P < 0.001).The treatment time by the RapidArc planning was 111.3 s,significantly longer than that by IMRT planning (103.6 s,t = 3.57,P = 0.002).Conclusions The RapidArc planning improves the dose distribution of CTV and homogeneity of PTV for breast cancer radiotherapy after breast-conserving surgery.However,it significantly enlarges the volume of normal tissues irradiated in low dose areas,prolongs the treatment delivery time,and increases the MU value in comparison with IMRT.

SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA