Chinese Journal of Medical Instrumentation ; (6): 240-245, 2021.
Artículo en Chino | WPRIM | ID: wpr-880459

RESUMEN

Aiming at the problem of timeliness of CBCT reconstruction, a CBCT fast short scan reconstruction method is proposed. At the same time, the image reconstruction process in which a new attenuation compensation algorithm is applied to improve image quality. When performing FDK three-dimensional reconstruction of a single-frame acquisition image, the Parker-weighted image is calculated in real time, and a new attenuation compensation algorithm is applied in the back projection process to complete the short scan Parker-weighted reconstruction. This method simulates the CBCT synchronous acquisition and reconstruction process by establishing collection and reconstruction threads. Under the premise of satisfying the reconstruction quality, the reconstruction can be completed within 1 to 2 seconds after the patient collection is completed, which achieves the purpose of real-time.


Asunto(s)
Humanos , Algoritmos , Tomografía Computarizada de Haz Cónico , Procesamiento de Imagen Asistido por Computador , Tomografía Computarizada de Haz Cónico Espiral