Artículo en Inglés | WPRIM | ID: wpr-100536

RESUMEN

BACKGROUND: Blood gas analysis plays a crucial role in critical care settings, and immediate and precise analysis improves clinical outcomes through prompt treatment. We evaluated the performance of a cartridge-type blood gas analyzer, i-Smart 300 (i-SENS, Korea), according to the Clinical and Laboratory Standard Institute (CLSI) guidelines and compared it to a conventional blood gas analyzer. METHODS: The precision was evaluated according to CLSI EP5-A3. The i-Smart 300 was compared to the Stat Profile Critical Care Xpress (STP CCX) (Nova CCX; Nova Biomedical, USA) according to CLSI EP9-A3 using the following eight parameters: pH, partial carbon dioxide pressure, partial oxygen pressure, sodium, potassium, chloride, ionized calcium, and hematocrit. Linearity was determined using five levels of control materials according to CLSI EP6-A. RESULTS: Within-run precision and total precision, demonstrated as coefficients of variation, ranged from 0.02 to 2.50% and from 0.05 to 3.46%, respectively. Correlation analysis yielded a correlation coefficient from 0.966 to 0.996 between the i-Smart 300 and the conventional analyzer (Nova CCX). The i-Smart 300 showed excellent linearity at eight parameters with acceptable percent recovery. CONCLUSIONS: The i-Smart 300, a portable cartridge-type blood gas analyzer, showed high precision and good correlation with a traditional bench-top blood gas analyzer. It could be useful in critical care settings.


Asunto(s)
Análisis de los Gases de la Sangre , Calcio , Dióxido de Carbono , Cuidados Críticos , Hematócrito , Concentración de Iones de Hidrógeno , Oxígeno , Presión Parcial , Sistemas de Atención de Punto , Potasio , Sodio
2.
Artículo en Coreano | WPRIM | ID: wpr-50882

RESUMEN

BACKGROUND: The i-Smart 30 point-of-care (POC) analyzer (i-SENS, Korea) is a compact and portable system used for the analysis of electrolytes (sodium, potassium, chloride) and hematocrit in whole blood samples. In this study, we evaluated the analytical performance of the i-Smart 30 analyzer. METHODS: Precision and sample-related percent carry-over were determined using the quality control materials. Comparison study was performed with the Stat Profile Critical Care Xpress (STP CCX; Nova Biomedical, USA) analyzer using venous whole blood samples. RESULTS: In the precision study, imprecision studies demonstrated within-run and total-run coefficients of variation within 0.5-3.9% and 0.7-4.4%, respectively, for all analytes. A good correlation was found between the i-Smart 30 analyzer and the STP CCX analyzer, except for chloride that showed high intercept. In the study of carry-over, sample-related carry-over for Na+, K+, Cl- and Hct were demonstrated as 0.84%, 0%, 0.86% and 1.56%, respectively. CONCLUSIONS: We conclude that the i-Smart 30 analyzer is suitable for routine use in clinical laboratories, especially where rapid test results are required such as emergency departments, intensive care units, and dialysis units. However, for Cl-, it is necessary that a significant correlation between this analyzer and a reference method should be demonstrated.


Asunto(s)
Cuidados Críticos , Diálisis , Electrólitos , Urgencias Médicas , Hematócrito , Unidades de Cuidados Intensivos , Sistemas de Atención de Punto , Potasio , Control de Calidad