Bangladesh Med Res Counc Bull ; 2007 Dec; 33(3): 92-7
Article Dans Anglais | IMSEAR | ID: sea-258

Résumé

Tissue phantom ratio (TPR), for square fields of various dimensions has been determined at varying depths in water. The dose in water has been measured at a fixed source-to-surface distance (SSD) of 100 cm and reference depth of 5 cm for 6 MV photon beam of Siemens Linear Accelerator Primus 11 in German Cancer Research Center (DKFZ), Heidelberg, Germany. A modified formula has been developed to calculate the TPR value for isocentric treatment. The present article describes the conversion of the measured data values into a comprehensive and consistent data set by the modified formula, that gives the TPR from Percentage Depth Dose (PDD) with depth as a function of field sizes from 10 mm x 10 mm upto 300 mm x 300 mm) and depth (from 0 mm to 300 mm).


Sujets)
Absorptiométrie photonique/instrumentation , Humains , Modèles théoriques , Fantômes en imagerie , Photons , Radiométrie/instrumentation