Asian Pac J Allergy Immunol ; 2004 Jun-Sep; 22(2-3): 109-14
Article Dans Anglais | IMSEAR | ID: sea-36711

Résumé

Skin prick testing (SPT) is widely used in the assessment of allergic disorders. Different SPT techniques are widely used. The aim of this study was to compare the response to SPT using a multiple lancet device (MLD) with the results of a single lancet (SL). Fifty patients with allergic rhinitis were included in this study. Initially, SPT was performed by a SL technique. After one week SPT was repeated using the MLD on all patients. The patients were tested with a panel containing 19 specific allergens including grass pollen, tree pollen, house dust mites, weed pollen allergen extracts, histamine and a negative control. The skin responses were recorded after 15 minutes for each device by measuring the diameter of the wheal and the erythema. The skin wheal responses for grass pollen, tree pollen, weed pollen and house dust mite allergen extracts obtained using the SL were generally significantly larger than those using the MLD. The comparison between the MLD and the SL methods revealed that SPT was positive with SL and negative with MLD in 176 tests (15.3%), and on the contrary SPT was positive with MLD and negative with SL in only 13 tests (1%). In conclusion, we claim that SPT using SL shows a higher degree of sensitivity and reproducibility.


Sujets)
Adulte , Femelle , Humains , Mâle , Rhinite/diagnostic , Sensibilité et spécificité , Tests cutanés/instrumentation