Journal of Southern Medical University ; (12): 1063-1065, 2010.
Article Dans Chinois | WPRIM | ID: wpr-289993

Résumé

<p><b>OBJECTIVE</b>To evaluate neointimal proliferation following placement of a new drug-eluting stent (BUMA) by optical coherence tomography (OCT).</p><p><b>METHODS</b>Twenty-two patients with coronary artery disease were randomized into BUMA group (n=15) and Endeavor group (n=7) and underwent OCT imaging after 9 months of stent implantation.</p><p><b>RESULTS</b>The neointima hyperplasia (NIH) thickness in BUMA group were significantly smaller than that in endeavor group (0.220-/+0.140 mm vs 0.269-/+0.207 mm, P<0.001), and the uncovered Struts were significantly lower in BUMA group than in Endeavor group (5.65% vs 6.56%, P<0.0001). The luminal late loss in BUMA group was also significantly lower (34.87-/+11.50 vs 40.82-/+18.53, P=0.025).</p><p><b>CONCLUSION</b>BUMA stent is safe and effective for treatment of coronary artery disease.</p>


Sujets)
Humains , Angioplastie coronaire par ballonnet , Prolifération cellulaire , Maladie des artères coronaires , Anatomopathologie , Thérapeutique , Vaisseaux coronaires , Anatomopathologie , Endoprothèses à élution de substances , Études prospectives , Tomographie par cohérence optique , Tunique intime , Anatomopathologie
2.
Chinese Journal of Cardiology ; (12): 585-589, 2009.
Article Dans Chinois | WPRIM | ID: wpr-236450

Résumé

<p><b>OBJECTIVE</b>To evaluate the characteristic of late stent malposition after drug-eluting stent implantation by optical coherence tomography (OCT).</p><p><b>METHODS</b>The study comprised of 32 patients (target vessels: 51, total stents: 71) underwent drug eluting stent implantation one year ago [average (14.8 +/- 5.2) months]. OCT images of the stent were analyzed at interval of 0.5 mm. The stent malposition was detected, the thickness of intima and gap between the stent strut and vessel wall were measured.</p><p><b>RESULTS</b>Stent malposition was detected in 7 patients without clinical cardiac events, including positive remodeling (n = 4), overlapping stents (n = 1) and stent struts covered with thrombus (n = 2). Stent strut apposition with complete intima coverage was about 97.6%, stent struts malposition was 2.4% including half of struts located at the ostium of side branch. The intima coverage of stent struts is similar between the struts at the side branch and others [(0.06 +/- 0.05) mm vs. (0.05 +/- 0.03) mm, P > 0.05].</p><p><b>CONCLUSION</b>The causes of late stent malposition include the primary malposition after stent implantation, positive remodeling, overlapping stents and stent struts located at the ostium of side branch. Thinner intima coverage was also observed on the stent struts with malposition.</p>


Sujets)
Sujet âgé , Femelle , Humains , Mâle , Adulte d'âge moyen , Endoprothèses à élution de substances , Tomographie par cohérence optique , Échec thérapeutique