Your browser doesn't support javascript.
loading
Montrer: 20 | 50 | 100
Résultats 1 - 1 de 1
Filtre
Ajouter des filtres








Gamme d'année
1.
China Pharmacist ; (12): 1850-1853, 2018.
Article Dans Chinois | WPRIM | ID: wpr-705725

Résumé

Objective: To establish a method for the determination of bilastine in human plasma, and evaluate the uncertainty by LC-MS/MS. Methods: The uncertainty sources were obtained from the whole process of the determination including repeatability, e-quipment error, weighting, solution preparation, calibration fitting and plasma sample handling. The uncertainty and synthesized un-certainty of each component were calculated, and then the expanded uncertainty was obtained. Results: The expanded uncertainty for low (15 ng·ml-1), medium (400 ng·ml-1) and high(1 200 ng·ml-1) level of bilastine was 1. 45 ng·ml-1, 28. 72 ng·ml-1 and 74. 61 ng·ml-1, respectively (k=2, P=95% ). Conclusion: The uncertainty in the determination of bilastine in human plasma is mainly caused by equipment error, solution preparation, protein precipitation and calibration fitting (especially at low level).

SÉLECTION CITATIONS
Détails de la recherche