Your browser doesn't support javascript.
loading
Montrer: 20 | 50 | 100
Résultats 1 - 1 de 1
Filtre
Ajouter des filtres








Gamme d'année
1.
Journal of Southern Medical University ; (12): 1352-1356, 2013.
Article Dans Chinois | WPRIM | ID: wpr-319413

Résumé

<p><b>OBJECTIVE</b>To propose a new method for acquiring pulmonary function parameters based on measurement of volume changes of thoracic impedance.</p><p><b>METHODS</b>We studied the relationship between the volume changes of thoracic impedance and pulmonary function parameters during forced breathing based on bioimpedance measurement, and developed an instrument for measuring thoracic impedance. Using this instrument and a MRI spirolab III lung function test instrument, both based on flowmeter measurement, we measured such pulmonary function parameters including forced vital capacity (FVC), forced expiratory volume in one second/FVC (FEV1/FVC), and peak expiratory flow in 10 healthy volunteers and compared the measurement results.</p><p><b>RESULTS</b>The differences in the parameters measured using the two instruments were all within two folds of the positive and negative standard deviations of the average values, demonstrating good consistency in the measurement between the two methods.</p><p><b>CONCLUSIONS</b>The measurement results of the bioimpedance-based instrument we developed show good consistency with those by the commercially available pulmonary function test instrument.</p>


Sujets)
Adulte , Humains , Jeune adulte , Impédance électrique , Volume expiratoire maximal par seconde , Débit expiratoire de pointe , Tests de la fonction respiratoire , Méthodes , Thorax , Physiologie , Capacité vitale