Chinese Journal of Interventional Cardiology ; (4): 261-265, 2017.
Article Dans Chinois | WPRIM | ID: wpr-609146

Résumé

Objective By comparing the efficacy and complication rates of the 8-mm-tip cryoablation catheter with the normal electrode ablation catheter in the treatment of atrioventricular nodal reentrant tachycardia,this study investigated the efficacy and feasibility of ablation with the 8-mm-tip cryoablation catheter.Methods This is a retrospective case-control study including 122 patients with AVNRT treated with CRYO (n =56) using an 8-mm-tip cryoablation catheter or RF ablation (n =66) from June 2014 to May 2016.The procedure success rate,the recurrence rate,atrioventricular block incidence,procedure time and the difference between the X-ray fluoroscopy dose were compared between the 2 groups.Results The procedure success rate was comparable between the 2 groups(100% for CRYO vs.98.5% for RF,P >0.999)and no AVB was found in both groups.The CRYO group needed shorter procedural time [(66.29±4.72)min vs.(70.00 ± 7.50) min,P =0.001] and less X-ray exposure [(674.14 ± 126.12) mSv vs.(837.52 ± 138.38) mSv,P > 0.001] than the RF group.Conclusions 8-mm-tip cryoablation catheter cryoablation for atrioventricular nodal reentrant tachycardia is as safe and effective as compared to conventional radiofrequency ablation with potential advantages.