International Eye Science ; (12): 646-650, 2024.
Article Dans Chinois | WPRIM | ID: wpr-1012837

Résumé

AIM: To assess the accuracy of predicting intraocular lens(IOL)power after myopic refractive surgery using the Pentacam system's true net power(TNP)in the 3 mm zone combined with the SRK/T formula [i.e. TNP 3 mm(SRK/T)].METHODS: Retrospective study. This study enrolled 35 cases(50 eyes)of patients undergoing cataract surgery after laser assisted in situ keratomileusis(LASIK)or photorefractive keratectomy(PRK)from July 2019 to December 2021. Preoperatively, IOL power of 50 eyes, 34 eyes and 41 eyes was calculated by TNP 3 mm(SRK/T), Barrett True-K and Olsen 2 formulas, respectively, with at least 2 formulas used to calculate IOL power for each patient. The actual diopter was recorded 3 mo postoperatively. Prediction errors(PE)of IOL power were compared among the three calculation methods, and the proportion of eyes with PE within ±0.5 D and ±1.0 D was analyzed.RESULTS: The PE at 3 mo postoperatively for TNP 3 mm(SRK/T), Barrett True-K, and Olsen 2 was -0.02±0.63, -0.54±0.80, and 0.25±0.80 D, respectively(P<0.001). The proportions of PE within ±0.5 D were 66%(33/50), 44%(15/34)and 37%(15/41), respectively(P<0.05); the proportions of PE within ±1.0 D were 88%(44/50), 71%(24/34)and 80%(33/41), respectively(P>0.05).CONCLUSION: The Pentacam TNP 3 mm(SRK/T)method is simple to operate and provides accurate calculation of IOL power after corneal refractive surgery.