Chinese Journal of Medical Instrumentation ; (6): 448-452, 2020.
Article Dans Chinois | WPRIM | ID: wpr-942760

Résumé

The purpose of this study was to investigate the NOAEL of the nickel ion and provide with basic data for the biological evaluation of those medical devices containing nickel. Five groups SD rats were repeatedly exposed during 14 d respectively to nickel at first stage doses of 4.9, 3.7, 2.5 mg/(kg.d), and the second stage doses of 1.2, 0.25 mg/(kg.d) by the intravenous route. The results showed that the NOAEL of nickel ion is 0.25 mg/(kg.d) for SD rats, and the result was verified by subchronic systemic toxicity test of nickel alloy. The threshold of toxicological concern (TTC) of nickel is 150 μg/d (based on application of 100-fold uncertainty factor and a body weight of 60 kg)deduced by these data.


Sujets)
Animaux , Rats , Équipement et fournitures/effets indésirables , Nickel/toxicité , Dose sans effet nocif observé , Rat Sprague-Dawley , Appréciation des risques