Your browser doesn't support javascript.
loading
Montrer: 20 | 50 | 100
Résultats 1 - 1 de 1
Filtre
Ajouter des filtres








Gamme d'année
1.
Chinese Journal of Medical Imaging Technology ; (12): 594-598, 2017.
Article Dans Chinois | WPRIM | ID: wpr-608676

Résumé

Objective To investigate the value of wide-detector Revolution CTA with 70 kV tube voltage and prospective ECG-gated technique in diagnosis of congenital heart disease (CHD) in infants and children.Methods Forty-five infants and children with complicated CHD underwent echocardiography and wide-detector Revolution CTA.According to the sur gical findings,the diagnostic efficiency of Revolution CTA and echocardiography were calculated and compared.The radiation effective dose (ED) and iodine dose were calculated.The quality of CT images was also evaluated.Results There were 25 separate cardiovascular anomalies including 6 congenital cardiac structure anomalies and 19 congenital extracardiac vascular anomalies.For congenital extracardiac vascular anomalies,there was significant difference of diagnostic accuracy and the detectable rate between CTA (99.77% [853/855],97.73% [86/88]) and echocardiography (98.71% [844/855],88.64% [78/88];x2 =6.28,5.72,both P<0.05).The average of ED was (0.20±0.05)mSv and the mean iodine dose was (2.06± 1.09)g.All CT images were qualified for diagnosis.Conclusion The wide-detector Revolution CTA,with the prospective ECG-gated technique and 70 kV tube voltage,can provide high accuracy for assessment of CHD in infants and children,which can keep good image quality,with the low radiation dose.

SÉLECTION CITATIONS
Détails de la recherche