Your browser doesn't support javascript.
loading
Montrer: 20 | 50 | 100
Résultats 1 - 1 de 1
Filtre
Ajouter des filtres








Gamme d'année
1.
Acta Pharmaceutica Sinica B ; (6): 482-486, 2015.
Article Dans Anglais | WPRIM | ID: wpr-310003

Résumé

A novel method for the simultaneous determination of 3-nitrotyrosine (NT) and 3-chlorotyrosine (CT) in human plasma has been developed based on direct analysis in real time-tandem mass spectrometry (DART-MS/MS). Analysis was performed in the positive ionization mode using multiple reaction monitoring (MRM) of the ion transitions at m/z 216.2/170.1 for CT, m/z 227.2/181.1 for NT and m/z 230.2/184.2 for the internal standard, d (3)-NT. The assay was linear in the ranges 0.5-100 μg/mL for CT and 4-100 μg/mL for NT with corresponding limits of detection of 0.2 and 2 μg/mL. Intra- and inter-day precisions and accuracies were respectively <15% and ±15%. Matrix effects were also evaluated. The method is potentially useful for high throughput analysis although sensitivity needs to be improved before it can be applied in clinical research.

SÉLECTION CITATIONS
Détails de la recherche