Your browser doesn't support javascript.
loading
Montrer: 20 | 50 | 100
Résultats 1 - 1 de 1
Filtre
Ajouter des filtres








Gamme d'année
1.
China Pharmacy ; (12): 2581-2582,2583, 2015.
Article Dans Chinois | WPRIM | ID: wpr-605109

Résumé

OBJECTIVE:To establish the purity determination of bisacodyl by differential scanning calorimetry(DSC)and the valuation of uncertainty. METHODS:DSC was conducted to detect the purity of bisacodyl and determine the optimal testing condi-tions. According to related standards,indium enthalpy change values,measurement repeatability,weighing process,instrument tem-perature deviation and system software deviation were systematically analyzed. The results were verified by HPLC. RESULTS:When the fiducial probability P was 0.95,the standard value and uncertainty of content of bisacodyl was (99.88 ± 0.06)% mea-sured by DSC. Weighing process,instrument temperature deviation and system software deviation had great effects on the total un-certainty. The result of HPLC and DSC were the same. CONCLUSIONS:The established DSC can quickly and accurately determine the chemical purity of bisacodyl. The uncertainty evaluation is reliable. Regularly calibrated and verificated equipment and strict con-trol of the weighing process will help to improve the accuracy measured by DSC;and it provides a new analysis method for the de-termination of purity of bisacodyl.

SÉLECTION CITATIONS
Détails de la recherche