Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Adicionar filtros








Intervalo de ano
1.
Military Medical Sciences ; (12): 55-59, 2016.
Artigo em Chinês | WPRIM | ID: wpr-491791

RESUMO

Objective To evaluate the performance of emergency test and the applicability under complex field conditions of biochemical modules of field point-of-care test ( POCT ) system ( type A ) .Methods The precision and anti-interference ability of albumin ( ALB ) , total bilirubin ( TBIL ) , alanine transaminase ( ALT ) , aspartate transaminase (AST),blood area nitrogen(BUN),creatinine(CREA),uric acid(UA),lactate dohydrogenase(LDH),creatine kinase ( CK) ,and glucose( GLU) detected by field POCT system( A) were analyzed according to standards formulated by National Committee for Clinical Laboratory Standards(NCCLS).Field POCT system(A) and Coulter Beckman AU2700 automatic biochemical analyzer were used to detect the serum of 22 clinical cases respectively.After simulating the field environment by adjusting the temperature and humidity, we compared the results of mixed serum under different environment conditions. Results The coefficients of variation in total precision of ALB,TBIL,ALT,AST,BUN,CREA,UA,LDH,CK,and GLU detected by field POCT system(A) were 3.34%,6.54%,6.01%,4.80%,3.95%,5.59%,3.33%,6.19%,7.40%,and 4.56%(LevelⅠ);and 3.08%,4.47%,4.02%,4.31%,3.76%,4.22%,2.93%,5.25%,6.39%,and 4.35%(LevelⅡ) respectively.When triglycerides( TG) level was at 21 mmol/L, the interference rate was below 10%.When bilirubin level was at 120 μmol/L, the interference rate of ALT,AST and CREA was -33.33%,-22.99%,20.00%(LevelⅠ), and -22.13%,-14.55%,and 8.70%(LevelⅡ),respectively.When its level was at 240 μmol/L, the interference rate of UA was -16.67%and -24.69%, respectively at two levels;if hemoglobin( Hb) was at 170 mg/dl, the interference rate of TBIL and LDH was 20.00%,and 99.26%(LevelⅠ),and 15.38%,and 40.79%(LevelⅡ),respectively;if it was at 340 mg/dl, the interference rate of ALT and AST was 9.84% and 13.79%(LevelⅠ), and 12.30%,and 12.27%(LevelⅡ),respectively;if it was at 510 mg/dl, the interference rate of CREA,UA and Ck in LevelⅠwas 26.67%, 16.67%,and 11.74%.The R2 of linear regression between field POCT system( A) and AU2700 automatic biochemical analyzer were 0.961,0.995,0.989,0.995,0.990,0.989,0.989,0.963,0.978,and 0.993, respectively.The POCT system could not work at 35℃ or higher temperature, and there was no difference in the results of detection between temperatures of 10-30℃or RH of 70%-90% and normal temperature and humidity(20℃,RH 50%) (P>0.05). However, the result of ALT and CK at high temperature and humidity was significantly higher than at normal temperature and humidity(P<0.001,and P=0.011, respectively).Conclusion The biochemical module of field POCT system(A) has a good correlation with the common large biochemical analyzer, and its precision meets the requirement of laboratory detection, but jaundice and hemolytsis can interfere in several tests to varying degrees.The POCT system can basically ensure accurate detection under field conditions of temperature and humidity, but should not work under extreme environments.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA