Indian J Ophthalmol ; 2013 Mar; 61(3): 129-131
Artigo em Inglês | IMSEAR | ID: sea-147884

RESUMO

The study was conducted to evaluate the intra-session repeatability of Tonopen AVIA (TPA). 180 eyes of 180 patients (50 eyes with glaucoma, 130 eyes of controls) were recruited for this observational study. The mean age of patients enrolled in the study was 43.9 ± 16.7 yrs (84 males, 96 females). Mean IOP recorded with Tonopen AVIA was 19.5 ± 9.5 mmHg, 19.4 ± 9.6 mmHg and 19.3 ± 9.2 mmHg, respectively in the first, second and third instances (P = 0.656). The intraclass correlation coefficient (ICC) ranged from 0.996 (95% CI: 0.956 - 0.998) for glaucoma subjects to 0.958 (95% CI: 0.934 - 0.975) for controls. The coefficient of variation in the study population ranged from 3.47% (glaucoma patients) to 8.10% (healthy controls), being 6.07% overall. The coefficient of repeatability varied between 2.96 (glaucoma patients), 3.35 (healthy controls) to 3.24 (overall). Thus, the Tonopen Avia shows good intrasessional repeatability of IOP in both glaucomatous patients and healthy subjects.