Journal of Biomedical Engineering ; (6): 1185-1191, 2023.
Artigo em Chinês | WPRIM | ID: wpr-1008949

RESUMO

A novel structural dynamics test method and device were designed to test the biomechanical effects of dynamic axial loading on knee cartilage and meniscus. Firstly, the maximum acceleration signal-to-noise ratio of the experimental device was calculated by applying axial dynamic load to the experimental device under unloaded condition with different force hammers. Then the experimental samples were divided into non-specimen group (no specimen loaded), sham specimen group (loaded with polypropylene samples) and bovine knee joint specimen group (loaded with bovine knee joint samples) for testing. The test results show that the experimental device and method can provide stable axial dynamic load, and the experimental results have good repeatability. The final results confirm that the dynamic characteristics of experimental samples can be distinguished effectively by this device. The experimental method proposed in this study provides a new way to further study the biomechanical mechanism of knee joint structural response under axial dynamic load.


Assuntos
Animais , Bovinos , Fenômenos Biomecânicos , Articulação do Joelho/fisiologia , Menisco , Fenômenos Mecânicos , Suporte de Carga