Biomedical and Environmental Sciences ; (12): 471-474, 2014.
Artigo em Inglês | WPRIM | ID: wpr-270578

RESUMO

To establish an analytical method for determination of four bisphenols (BPA, BPB, BPF, and BPS) and two alkylphenols (4-n-OP, 4-n-NP) in water by ultra performance liquid chromatography- tandem mass spectrometry (UPLC/MS/MS). The water samples were extracted and condensed with solid-phase extraction (SPE) using C18 cartridges and eluted by acetonitrile. Separation was carried out with Acquity BEH C8 column and detection were performed by UPLC/MS/MS. Quantification was calculated by using the internal standard BPA-d16 and 4-n-NP-d8. The linear correlation coefficients of these compounds in the range of 1.0-100.0 μg/L were all over 0.999. The minimum detectable concentrations were 0.75-1.0 ng/L, and the recoveries ranged from 87.0% to 106.9%. Relative standard deviations (RSDs) were between 1.26% and 3.67%. Applying this method to detect the source water of Chaohu Lake and drinking water of Hefei, six target compounds were detected in different levels. This method is simple with high sensitivity and selectivity, could be suitable for the determination of these compounds in source and drinking water.


Assuntos
Cromatografia Líquida , Disruptores Endócrinos , Fenóis , Extração em Fase Sólida , Espectrometria de Massas em Tandem , Poluentes Químicos da Água