Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Adicionar filtros








Intervalo de ano
1.
Chinese Journal of Practical Nursing ; (36): 1608-1612, 2023.
Artigo em Chinês | WPRIM | ID: wpr-990380

RESUMO

Objective:To design a tracheotomy cannula cuff filling device for hyperbaric oxygen therapy, which is convenient for clinical operation, improves work efficiency and reduces the incidence of aspiration pneumonia.Methods:This study was a randomized controlled trial. From July 2020 to June 2022, 90 patients with tracheotomy who were treated with hyperbaric oxygen in the First Hospital of Jiaxing were selected as the research objects. According to the random number table method, the patients were divided into the experimental group and the control group, with 45 cases in each group. In the experimental group, the cuff pressure was maintained by the tracheotomy cannula cuff filling device, and in the control group, the traditional water injection method was used to maintain the cuff pressure. The operation time, infection index and incidence of aspiration pneumonia were compared between the two groups.Results:The operation time in the experimental group was (6.33 ± 1.31) s lower than that in the control group (40.96 ± 3.70) s, and the difference was statistically significant ( t=-59.11, P<0.05). Body temperature, C-reactive protein and procalcitonin after treatment in the experimental group were (36.91 ± 0.83) ℃, (34.59 ± 16.25) mg/L, (1.57 ± 0.82) μg/L, respectively, lower than those in the control group (37.42 ± 0.72) ℃, (44.18 ± 18.10) mg/L, (2.45 ± 0.92) μg/L, the differences were statistically significant ( t=-3.09, -2.64, -4.73, all P<0.05). The difference of white blood cell count post-treatment between the two groups was not statistically significant ( P>0.05). The incidence of aspiration pneumonia in the experimental group was 11.11%(5/45) lower than 31.11%(14/45) in the control group, and the difference was statistically significant ( χ2=5.17, P<0.05). Conclusions:The application of tracheotomy cannula cuff filling device can simplify the operation process, reduce the incidence of infection and aspiration pneumonia, and optimize the clinical work.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA