Chinese Journal of Laboratory Medicine ; (12)2001.
Artigo em Chinês | WPRIM | ID: wpr-685370

RESUMO

Objective To find a proper way for accurate results on completing blood counts.Methods Based on the results from automatic hematology analyzer XE-2100,set up the criteria for blood cell microscopic examination.1368 blood specimen were detected and the results were analyzed according to the criteria.Statistics on the data were made to evaluate the accordance between warnings of analyzer and manual examination,likewise the reliability of the criteria.Results Comparing with microscopic examination,analyzer warning on low PLT has good accordance,Kappa value was 0.95,U value was 35.19,P