Al-Shifa Journal of Ophthalmology. 2011; 7 (1): 39-46
em Inglês | IMEMR | ID: emr-130259

RESUMO

To determine the validity of frequency doubling technology [FDT] in identifying glaucoma subjects using Humphrey perimetry as gold standard. Cross-sectional validation study. The study was conducted at the Glaucoma clinic of Al-Shifa Trust Eye Hospital Rawalpindi from 16[th] Nov 2009 to 15[th] May 2010 on 84 eyes suspected to have glaucoma. After an informed consent, detailed history and complete ocular examination, patients were briefed about the two diagnostic tests. Each patient performed both perimetry tests, i.e. Standard Automated Perimetry [Humphrey 30-2] and FDT. At the end of each test, perimetry systems calculated mean deviation for the eye examined. Sensitivity and specificity of the procedures was calculated using formulae. Mean test duration for FDT perimetry was 4.44 min and for Humphrey perimetry duration was 9.10 min. The sensitivity and specificity calculated for FDT perimetry were 85.4% and 90.7% respectively. Area under the receiver operating curve for FDT perimetry was 0.880 [95%CI=0.799-0.961]. FDT perimetry has ability to discriminate between glaucomatous and healthy eyes, in less duration and can be effectively used as a screening tool to diagnose glaucoma


Assuntos
Humanos , Feminino , Masculino , Reprodutibilidade dos Testes , Estudos Transversais , Testes de Campo Visual , Técnicas e Procedimentos Diagnósticos