Ann Card Anaesth ; 2016 Apr; 19(2): 347-350
Artigo em Inglês | IMSEAR | ID: sea-177409
3.
Ann Card Anaesth ; 2012 Jan; 15(1): 47-49
Artigo em Inglês | IMSEAR | ID: sea-139634

RESUMO

Interference of monitored electrocardiogram (ECG) is a common event in intensive care units and operation theaters. Artifacts in the ECG tracing can resemble serious arrhythmia, thus leading to unnecessary usage of antiarrhythmics or electrical defibrillation. In addition, ECG artifacts may lead to serious hemodynamic consequences secondary to intra-aortic balloon pump (IABP) trigger failure. We report a rare event of IABP failure due to ECG artifact, which appeared on placement of the transthoracic echocardiography probe over the chest. Subsequent evaluation revealed potential current leakage from echocardiography machine secondary to earthing defect in the machine.


Assuntos
Artefatos , Eletrocardiografia , Hemodinâmica , Humanos , Balão Intra-Aórtico , Masculino , Pessoa de Meia-Idade