Chinese Journal of Nosocomiology ; (24)2005.
Artigo em Chinês | WPRIM | ID: wpr-590276

RESUMO

OBJECTIVE To analyze the monitoring performance and testing efficacy and reliability of standardized disposable B-D test pack and standard linen test pack,and summarize the corrective actions based on failure analyses for future reference.METHODS The two different test packs were employed to evaluate the vacuum performance of dynamic air removal sterilizer.RESULTS We enrolled 400 standardized disposable B-D test packs and standard linen test packs respectively.The disposable pack failed 4 cases with the success ratio 99%,while the linen pack failed 30 cases with the success ratio 92.5%.CONCLUSIONS The poor conformity of hand-made linen test pack,variation of steam pressure,sterilizer failure,and unprofessionalism of sterilizer operator contribute as main failure causes in B-D test.Standardized disposable B-D test pack can decrease subjective factors significantly,and the test results are more reliable and standardized.