1.
Chinese Journal of Digestive Endoscopy ; (12): 628-630, 2012.
Artigo em Chinês | WPRIM | ID: wpr-421012

RESUMO

ObjectiveTo evaluate the influence of electrotome on permanent and temporary cardiac pacemaker in laparoscopic cholecystectomy ( LC),and the application of cardiac pacemaker to the cases of cholecystolithiasis combined with bradyarrhythmia.MethodsClinical data of 215 patients with permanent or temporary cardiac pacemaker who underwent were studied for the preoperative and postoperative variation of pacemaker function,and for the influence of electricity coagulation during the operation on cardiac pacemaker function.ResultsLC was successfully completed in all 215 patients.The function of cardiac pacemaker was not obviously interfered during the operation,and the parameters of cardiac pacemaker did not remarkably change after the operation.ConclusionCardiac pacemaker is slightly interfered when electrotome and electrocoagulation were used in LC; LC is feasible and safe for patients with bradyarrhythmia by placement of cardiac pacemaker.