Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Adicionar filtros








Intervalo de ano
1.
Int J Pharm Pharm Sci ; 2019 May; 11(5): 37-42
Artigo | IMSEAR | ID: sea-205895

RESUMO

Objective: To develop and validate simple, sensitive stability indicating HPTLC (High performance thin layer chromatography) method for apixaban. Methods: The chromatographic separation was performed on aluminium plates precoated with silica gel 60 F254 using toluene: ethyl acetate: methanol (3:6:1 v/v/v) as mobile phase followed by densitometric scanning at 279 nm. Results: The chromatographic condition shows sharp peak of apixaban at Rf value of 0.38±0.03. Stress testing was carried out according to international conference on harmonization (ICH)Q1A (R2) guidelines and the method was validated as per ICH Q2(R1) guidelines. The calibration curve was found to be linear in the concentration range of 100-500 ng/band for apixaban. The limit of detection and quantification was found to be 11.66ng/bandand35.33ng/band, respectively. Conclusion: A new simple, sensitive, stability indicating high performance thin layer chromatographic (HPTLC) method has been developed and validated for the determination of apixaban.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA