International Eye Science ; (12): 846-849, 2015.
Artigo em Chinês | WPRIM | ID: wpr-637292

RESUMO

?AlM: To investigate repeatability and accuracy of a latest Keratograph for evaluating the tear film stability and to compare its measurements with that of traditional examination methods. ?METHODS: The results of noninvasive tear film break-up time ( Nl-BUT ) including the first tear film break-up time ( BUT-f ) and the average tear film break-up time ( BUT - ave ) were measured by Keratograph. The repeatability of the measurements was evaluated by coefficient of variation ( CV ) and intraclass correlation coefficient ( lCC) . Wilcoxon Signed-Rank test was used to compare Nl-BUT with fluorescein tear film break-up time ( FBUT) to confirm the correlation between Nl-BUT and FBUT, Schirmer l test values. Bland-Altman analysis was used to evaluate consistency. ?RESULTS: The study recruited 48 subjects ( 48 eyes ) (mean age 38. 7±15. 2 years). The CV and lCC of BUT-f were respectively 12. 6% and 0. 95, those of BUT-ave were 9. 8% and 0. 96. The value of BUT-f was lower than that of FBUT. The difference had statistical significance ( 6. 16±2. 46s vs 7. 46±1. 92s, P ?CONCLUSlON: Keratograph can provide Nl-BUT data that has a better repeatability and reliability, which has great application prospects in diagnosis and treatment of dry eye and refractive corneal surgery.