Chinese Journal of Radiological Health ; (6): 61-67, 2024.
Artigo em Chinês | WPRIM | ID: wpr-1012772

RESUMO

Objective To provide a theoretical basis for radiation health supervision through an analysis of the situation of computed tomography (CT) equipment quality control and CT room radiological protection in Guangdong Province, China in recent years. Methods We collected the data of 392 times of CT quality control and radiological protection testing by a third-party radiological health technical service institution in Guangdong Province from 2019 to 2021. We analyzed the levels of CT-owning hospitals, CT manufacturers, CT quality control test results, and the pass rate of radiation protection tests. Results The examined CT scanners were from different levels of hospitals in Guangdong Province, and were manufactured by nine major CT equipment manufacturers at home and abroad. The pass rate of CT room radiological protection was 99.88%, and the ambient dose equivalent rates of five monitoring points exceeded the limit, with four at the control room door and one at the shield wall of the room. The overall pass rate of CT equipment quality control was 99.49%, and the non-conforming parameters were the accuracy of positioning light and the deviation of reconstructed slice thickness. Conclusion In recent years, CT equipment quality control and room radiation protection in Guangdong Province have been at a high level.