RESUMO
Objective To analyze the status of quality indicators(QI) on specimen acceptability and establish preliminary qual ity specification.Methods Web based External Quality Assessment system was used to collect data of laboratories partici pated in "Medical quality control indicators in clinical laboratory" from 2015 to 2017,including once in 2015 and 2017 and twice in 2016.Rate and sigma scales were used to evaluate incorrect sample type,incorrect sample container,incorrect fill level and anticoagulant sample clotted.The 25th percentile (P25) and 75th percentile (P75) of the distribution of each QI were employed to establish the high,medium and low specification.Results 5 346,7 593,5 950 and 6 874 laboratories sub mitted the survey results respectively.The P50 of biochemistry (except incorrect fill level),immunology and microbiology reach to 6σ.The P50 of clinical laboratory is 4 to 6σ except for incorrect sample container.There is no significant change of the continuous survey results.Based on results in 2017 to establish the quality specification,the P25 and P75 of the four QIs is 0 and 0.084 4 %,0 and 0.047 6 %,0 and 0.114 2 %,0 and 0.078 4 %,respectively.Conclusion According to the results of the survey,most laboratories had a faire performance in biochemistry,immunology and microbiology,and clinical laboratory needs to be strengthened.Laboratories should strengthen the laboratory information system construction to ensure the actual and reliable data collection,and make a long time monitoring to achieve a better quality.