China Medical Equipment ; (12): 29-31, 2017.
Artigo em Chinês | WPRIM | ID: wpr-611399

RESUMO

Objective:To discuss the quality control method of injection pump and analyze the factors that affect the quality control of injection pump.Methods: Flow error and blocking alarm error were tested by the INFUTEST 2000E dual-channel infusion pump analyzer according to the national calibration specifications.Results: The syringe specifications, injection speed and injection solvent could influence the test for accuracy of injection pump. The lower injection speed set, the bigger flow test error was, while the smaller the blocking alarm test error was. On the other hand, the higher the injection speed set, the smaller the flow test error was and the bigger the blocking alarm test error was. Besides, the higher qualified rate of flow test was, the lower the qualified rate of blocking alarm test was.Conclusion: The quality control of injection pump should abide by the principle of controllable risk, and should not pursuit too much accuracy.

2.
Artigo em Chinês | WPRIM | ID: wpr-497740

RESUMO

Objective To investigate the maternal attitude and demands for Institute of Postpartum Care in order to provide targeted services.Methods 285 women were a investigated about maternal attitude and demands towards Institute of Postpartum Care with self-designed questionnaire.Results 97.9% (279/285) women thought sitting the month (puerperal period) was very important,50.2% (143/285)women were interested in Institute of Postpartum Care.They thought the main three advantages were energy saving,medical expertise and technology and avoidance of family conflicts;The three major issues were high price,lack of monitoring mechanism and unfamiliar environment and less family company;33 items of services demands over four points accounted for 78.6%,the three modules with highest scores were cleaning services scored 4.64±0.47,newborn services scored 4.45±0.41,consultation service scored 4.41:±0.53.Conclusions Professional postnatal care services have been a trend and social need,and a large proportion of women have great interest and demand on Institute of Postpartum Care which should make full use of its concentrate resources,and provide specific services according to maternal demands for services.It is urgent to build unified service standards and evaluation indicators to improve service quality.

3.
Artigo em Chinês | WPRIM | ID: wpr-482465

RESUMO

To master the performances of the cardiac defibrillator-monitor and enhance its reliability and effi-ciency through quality control. ESA612 electrical safety analyzer and Phase3 defibrillator analyzer were used to test the performances of 56 cardiac defibrillator-monitors, involving in basic items detection, electrical safety detection and performance detection. The qualified items included charging times, internal discharge, ECG monitoring and etc, while the unqualified ones included appearance, electrical safety as well as delay time under synchronous mode. The quality control can improve the safety of the cardiac defibrillator-monitor.

4.
China Medical Equipment ; (12): 116-117, 2014.
Artigo em Chinês | WPRIM | ID: wpr-459328

RESUMO

Objective: To apply the norms of the portable blood glucose meter in clinic and improve the quality and level of blood glucose detection. Methods:The management organization of the portable blood glucose meter was established. The organization understudied the working principles and the advantages and disadvantages of blood glucose meter. The model of the portable blood glucose meter was unified in a medical unit. The performance of the portable blood glucose meter and the biochemical analyzer was compared. The organization established standard operating procedures, strengthen personnel training. The clinical staff takes quality control regular. Results:The qualified rate of the use, the maintenance and the results detection was increased. Conclusion:The standardized management system guarantees the detection result of the portable blood glucose meter.