Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Adicionar filtros








Intervalo de ano
1.
Korean Journal of Blood Transfusion ; : 13-19, 2012.
Artigo em Inglês | WPRIM | ID: wpr-76698

RESUMO

BACKGROUND: Use of universal leukoreduction for prevention of leukocyte associated transfusion reactions is common practice in many countries. This study was conducted in order to evaluate the performance of a newly developed leukoreduction filter for red blood cells (RBCs), the RF300 (Kolon Industries, Inc, Gumi, Korea). METHODS: Filtration time, RBC recovery, residual leukocyte count, and leukocyte removal rate were evaluated. To assess the quality of RBCs after filtration, percent hemolysis was monitored for a period of 21 days. Performance of the RF300 (N=78) was compared with that of the Bio-R O2 plus (Fresenius, Hamburg, Germany), the Pall Purecell RC (Pall Co., Washington, USA), and the Sepacell R-500N (Asahi, Tokyo, Japan). RESULTS: The shortest filtration time was observed using the RF300 (P<0.05). Using the RF300, recovery of RBC was 96.5%, which was higher than that of two filters (P<0.05). Mean residual leukocyte count was 0.26x10(6)/unit, with a leukocyte removal rate of 3 log. Using the RF300, mean percent hemolysis was 0.32% at day 21, which was comparable with that of two filters, but lower than that of one filter (P<0.05). CONCLUSION: The RF300 meets all established quality requirements for conduct of safe and effective leukoreduction of RBCs.


Assuntos
Incompatibilidade de Grupos Sanguíneos , Colódio , Eritrócitos , Filtração , Hemólise , Contagem de Leucócitos , Leucócitos , Tóquio , Washington