Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Adicionar filtros








Intervalo de ano
1.
Chinese Journal of Laboratory Medicine ; (12)2003.
Artigo em Chinês | WPRIM | ID: wpr-685563

RESUMO

Objective To design an experiment for method-comparison and bias estimation of multi tests on multi instruments.Methods According to the procedure described by the NCCLS approved guideline EP9-A and improving the method of sample collection,we took 8 patient mixing samples each day to analyze all comparison tests on 11 auto-chemistry analyzer within following 5 days.The duplicates were assessed within the same run.The coefficient of correlation and average bias% were calculated,and the system errors at medical decided levels were assessed.Results Taking ALT as an example,the coefficients of correlation were between 0.994-1.000,and the average bias% were between-0.460%-4.927%,SE at 40U/L was-1.510-1.834 and SE at 300 U/L was-3.101-9.188.Conclusion In all tests that joined the comparison among the different instruments,28 tests were acceptable,2 tests were acceptable after modifying the coefficients,and AMY and LIP were not acceptable.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA