Artigo em Chinês | WPRIM | ID: wpr-708059

RESUMO

Objective To use thermoluminescense dosimeters (TLDs) to evaluate the radiation doses to various critical organs in the computerized imaging reference systems (CIRS) 5 years old pediatric anthropomorphic phantom result ing from Varian kilovohage cone beam CT (kV-CBCT) system based on the standard scanning protocols.Effective dose were also calculated based on dose measurement.Methods A batch of TLDs with consistency no larger than 2% were selected and annealed.First,the doses in an anthropomorphic pelvic phantom were measured using a CT chamber and TLDs,respectively,based on the standard pelvic protocols.The ratio of the both measurements is the TLD conversion coefficient.Other TLDs from the same batch were placed between two tissue-equivalent inserts and placed into the pre-drilled organ cavities of the pediatric phantom.By using standard protocols,the organs dose were measured,based on which the corresponding effective doses were calculated.Results The TLD conversion coefficient was 3.91 mGy/per reading.By using the standard head,low-dose thorax,pelvis protocol,the whole body effective dose was 0.63,6.85 and 19.3 mSv,respectively.Conclusions It is feasible for using the CT chamber-calibrated TLDs to measure the radiation doses from kV-CBCT to pediatric anthropomorphic phantom.The effective dose in pelvic protocol was higher than in thorax and head protocol,indicating that the pelvic protocol has a penitential to lead to larger radiation damage and higher risk of secondary cancer.