Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Adicionar filtros








Intervalo de ano
1.
Journal of Laboratory Medicine and Quality Assurance ; : 189-194, 2008.
Artigo em Coreano | WPRIM | ID: wpr-130598

RESUMO

BACKGROUND: CELL-DYN Sapphire (Abbott Diagnostics, Santa Clara, CA, USA) hematology analyzer is based on laser multi angle polarized scatter separation, focused flow impedance and fluorescent flow cytometry. In order to assess the usefulness of this newly introduced analyzer, we evaluated the precision and correlation of complete blood count including white blood cell differential count and reticulocyte count of CELL-DYN Sapphire instrument. METHODS: Patient samples ordered for complete blood count and white blood cell differential count and CELL-DYN 29 Plus control samples were used. We evaluated the precision of complete blood count and WBC differential count and analyzed correlation of these parameters for comparison with ADVIA 120 (Bayer corporation, Tarrytown, NY, USA) and manual count. We additionally evaluated reticulocyte parameters for precision and comparison with ADVIA 120. RESULTS: Short term precisions showed low coefficient of variation (CV) values for all CBC parameters including lower than 3% CV values for neutrophil and lymphocyte count. Long term precision showed lower than 4% CV values for all CBC and WBC differential count except for basophils. Comparison with ADVIA 120 showed good correlation for CBC except for MCHC. Good correlation was confirmed in all differential counts except for basophils. Comparison with manual WBC differential count showed good correlation except for monocyte and eosinophil. CONCLUSIONS: CELL-DYN Sapphire has high precision and good correlation with ADVIA 120 and manual count. CELL-DYN Sapphire is an exceptional instrument for CBC with WBC differentials and reticulocyte counting.


Assuntos
Humanos , Óxido de Alumínio , Basófilos , Contagem de Células Sanguíneas , Impedância Elétrica , Citometria de Fluxo , Hematologia , Leucócitos , Contagem de Linfócitos , Monócitos , Neutrófilos , Contagem de Reticulócitos , Reticulócitos
2.
Journal of Laboratory Medicine and Quality Assurance ; : 189-194, 2008.
Artigo em Coreano | WPRIM | ID: wpr-130591

RESUMO

BACKGROUND: CELL-DYN Sapphire (Abbott Diagnostics, Santa Clara, CA, USA) hematology analyzer is based on laser multi angle polarized scatter separation, focused flow impedance and fluorescent flow cytometry. In order to assess the usefulness of this newly introduced analyzer, we evaluated the precision and correlation of complete blood count including white blood cell differential count and reticulocyte count of CELL-DYN Sapphire instrument. METHODS: Patient samples ordered for complete blood count and white blood cell differential count and CELL-DYN 29 Plus control samples were used. We evaluated the precision of complete blood count and WBC differential count and analyzed correlation of these parameters for comparison with ADVIA 120 (Bayer corporation, Tarrytown, NY, USA) and manual count. We additionally evaluated reticulocyte parameters for precision and comparison with ADVIA 120. RESULTS: Short term precisions showed low coefficient of variation (CV) values for all CBC parameters including lower than 3% CV values for neutrophil and lymphocyte count. Long term precision showed lower than 4% CV values for all CBC and WBC differential count except for basophils. Comparison with ADVIA 120 showed good correlation for CBC except for MCHC. Good correlation was confirmed in all differential counts except for basophils. Comparison with manual WBC differential count showed good correlation except for monocyte and eosinophil. CONCLUSIONS: CELL-DYN Sapphire has high precision and good correlation with ADVIA 120 and manual count. CELL-DYN Sapphire is an exceptional instrument for CBC with WBC differentials and reticulocyte counting.


Assuntos
Humanos , Óxido de Alumínio , Basófilos , Contagem de Células Sanguíneas , Impedância Elétrica , Citometria de Fluxo , Hematologia , Leucócitos , Contagem de Linfócitos , Monócitos , Neutrófilos , Contagem de Reticulócitos , Reticulócitos
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA