Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Adicionar filtros








Intervalo de ano
1.
Annals of Laboratory Medicine ; : 319-325, 2013.
Artigo em Inglês | WPRIM | ID: wpr-178349

RESUMO

BACKGROUND: The urinary iodine micromethod (UIMM) is a modification of the conventional method and its performance needs evaluation. METHODS: UIMM performance was evaluated using the method validation and 2008 Iodine Deficiency Disorders survey data obtained from four urinary iodine (UI) laboratories. Method acceptability tests and Sigma quality metrics were determined using total allowable errors (TEas) set by two external quality assurance (EQA) providers. RESULTS: UIMM obeyed various method acceptability test criteria with some discrepancies at low concentrations. Method validation data calculated against the UI Quality Program (TUIQP) TEas showed that the Sigma metrics were at 2.75, 1.80, and 3.80 for 51+/-15.50 microg/L, 108+/-32.40 microg/L, and 149+/-38.60 microg/L UI, respectively. External quality control (EQC) data showed that the performance of the laboratories was within Sigma metrics of 0.85-1.12, 1.57-4.36, and 1.46-4.98 at 46.91+/-7.05 microg/L, 135.14+/-13.53 microg/L, and 238.58+/-17.90 microg/L, respectively. No laboratory showed a calculated total error (TEcalc)

Assuntos
Humanos , Iodo/urina , Laboratórios/normas , Controle de Qualidade , Espectrofotometria/normas , Urinálise/normas
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA