Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 4 de 4
Filtrar
Adicionar filtros








Intervalo de ano
1.
The Journal of Advanced Prosthodontics ; : 102-106, 2009.
Artigo em Inglês | WPRIM | ID: wpr-107293

RESUMO

STATEMENT OF PROBLEM: Screw loosening has been a common complication and still reported frequently. PURPOSE: The purpose of this study was to evaluate abrasion of the implant fixture and TiN coated abutment screw after repeated delivery and removal with universal measuring microscope. MATERIAL AND METHODS: Implant systems used for this study were Osstem and 3i. Seven pairs of implant fixtures, abutments and abutment screws for each system were selected and all the fixtures were perpendicularly mounted in liquid unsaturated polyesther with dental surveyor. After 20 times of repeated closing and opening test, the evaluation for the change of inner surface of implant and TiN-coated abutment screw, and weight loss were measured. Mann-Whitney test with SPSS statistical software for Window was applied to analyze the measurement of weight loss. RESULTS: TiN-coated abutment screws of Osstem and 3i showed lesser loss of weight than non-coated those of Osstem and 3i (P < .05, Mann-Whitney test). CONCLUSION: Conclusively, TiN coating of abutment screw showed better resistance to abrasion than titanium abutment screw. It was concluded that TiN coating of abutment screw would reduce the loss of preload with good abrasion resistance and low coefficient of friction, and help to maintain screw joint stability.


Assuntos
Colódio , Fricção , Articulações , Estanho , Titânio , Redução de Peso
2.
The Journal of Korean Academy of Prosthodontics ; : 601-610, 2007.
Artigo em Coreano | WPRIM | ID: wpr-179761

RESUMO

STATEMENT OF PROBLEM: The adhesion between titanium and ceramic is less optimal than conventional metal-ceramic bonding, due to reaction layer form on cast titanium surface during porcelain firing. PURPOSE: This study characterized the effect of titanium-ceramic adhesion after gold and TiN coating on cast and wrought titanium substrates. MATERIAL AND METHOD: Six groups of ASTM grade II commercially pure titanium and cast titanium specimens(13mm x 13mm x 1mm) were prepared(n=8). The conventional Au-Pd-In alloy served as the control. All specimens were sandblasted with 110micrometer Al2O3 particles and ultrasonically cleaned for 5min in deionized water, and dried in air before porcelain firing. An ultra-low-fusing dental porcelain(Vita Titankeramik) was fused on titanium surfaces. Porcelain was debonded by a biaxial flexure test at a cross head speed of 0.25mm/min. The excellent titanium-ceramic adherence was exhibited by the presence of a dentin porcelain layer on the specimen surface after the biaxial flexure test. Area fraction of adherent porcelain(AFAP) was determined by SEM/EDS. Numerical results were statistically analyzed by one-way ANOVA and Student- Newman-Keuls test at alpha=0.05. RESULTS: The AFAP value of cast titanium was greatest in the group 2 with TiN coating, followed by group 1 with Au coating and the group 3 with Al2O3 sandblasting. Significant statistical difference was found between the group 1, 2 and the group 3 (p.05). The AFAP values of the cast titanium and the wrought titanium were similar. However the group treated with Al2O3 sandblasting showed significantly lower value (p<.05).


Assuntos
Ligas , Cerâmica , Porcelana Dentária , Dentina , Incêndios , Cabeça , Estanho , Titânio , Água
3.
The Journal of Korean Academy of Prosthodontics ; : 119-130, 2007.
Artigo em Coreano | WPRIM | ID: wpr-218244

RESUMO

STATEMENT OF PROBLEM: A few dry lubricants were applied to abutment screws for the improvement of joint stability. PURPOSE: The purpose of this study was to evaluate the surface change and fit of TiNCoated abutment screw through the examination of tested screws in the field emission scanning electron microscope(FE-SEM;Netherland, Phillips co., model:XL 30 SFEG) after repeated closing and opening. MATERIALS AND METHOD: Titanium(3i/implant Innovations Inc, USA) and Gold-Tite abutment screws(3i/implant Innovations Inc, USA) were selected for Group A and C respectively. TiN coated titanium abutment screws were also divided into two groups, Group B and D. Abutment screws of each group and the fit of abutment screw/implant fixture/abutment were observed on FE-SEM after repeated closing and opening test respectively. RESULTS: 1. The abutment screws of TiN coated groups(Group B and D) showed more remarkable wear resistance in the threads of the screw than those of the other group(Group A and C). 2. There were more severe wear and defect of TiN coating in Group D tightened to 32 Ncm than Group B to 20 Ncm. 3. There was no difference in the fit of abutment screw/fixture/abutment among four groups, Group A, Group B, Group C and Group D. CONCLUSION: Under the conditions of this study, it is suggested that TiN coating of abutment screw be clinically acceptable and be expected to reduce the risk of abutment screw loosening. TiN coating of abutment screw showed good resistance against wear and the adequate fit of abutment screw/implant fixture/abutment.


Assuntos
Articulações , Lubrificantes , Estanho , Titânio
4.
The Journal of Korean Academy of Prosthodontics ; : 329-338, 2007.
Artigo em Inglês | WPRIM | ID: wpr-25863

RESUMO

PURPOSE: The aim of this study is to evaluate the effect of TiN coating of abutment screw on the unscrewing torque. MATERIAL AND METHODS: Titanium and Gold-Tite abutment screws were classified into two groups, Group A and C respectively, as control groups. Titanium abutment screws with TiN coatings were also classified into two groups, Group B and D, as experimental ones. Group A and B were tightened to 20 Ncm input torque, and Group C and D were tightened to 32 Ncm torque. Detorque values were measured with digital torque gauge during repeated closing and opening experiment. RESULTS: Abutment screws with TiN coating (Group B and D) showed statistically significant higher mean detorque values than those of Group A and C. DISSCUSION: Physical properties of TiN coating, such as low friction coefficient, high hardness and wear resistance, might contribute to higher detorque values. CONCLUSION: It is suggested that TiN coating of abutment screw help to reduce the risk of screw loosening and improve the stability of screw joint.


Assuntos
Fricção , Dureza , Articulações , Estanho , Titânio , Torque
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA