Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Adicionar filtros








Intervalo de ano
1.
The Journal of the Korean Orthopaedic Association ; : 373-376, 2001.
Artigo em Coreano | WPRIM | ID: wpr-645161

RESUMO

PURPOSE: The purpose of this study was to compare the two-dimensional with three-dimensional radiographic measurements of polyethylene liner wear in the Harris-Galante II acetabular cup. MATERIALS AND METHODS: We measured the polyethylene liner wear amount and the wear rate of 64 hips with Harris-Galante II acetabular cup by two methods: Livermore's method and modified Devane's method. We evaluated the relationships between the two methods. RESULTS: The average total amount of wear and the average wear rate are 0.86+/-1.01 mm and 0.18+/-0.20 mm/year using the two-dimensional method and 0.99+/-1.01 mm and 0.21+/-0.2 mm/year using three-dimensional method respectively. The amount of wear and the wear rate by the three-dimensional method was larger than those by the two dimensional method. CONCLUSION: The amount of polyethylene liner wear can be measured by the two-dimensional method because it is highly correlated with the amount that is measured by the three-dimensional method, since there was no differences between the two methods.


Assuntos
Acetábulo , Quadril , Polietileno
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA