Korean Journal of Ophthalmology ; : 154-160, 2004.
Artigo em Inglês | WPRIM | ID: wpr-94532

RESUMO

Existing methods of stereoacuity testing need specific glasses or optical device for use. We have designed a new stereoacuity test for the digitalized, random-dot stereogram and researched its clinical usefulness. A digitalized, random-dot, stereoacuity test card was created with a computer program that used a preferred symbol and the designed system was tested along with the Randot preschool stereoacuity, Titmus-fly and Lang tests to compare their sensitivity and specificity. The mean success rate of the digitalized, random-dot test was 98.2%, while the rates of the Randot preschool stereoacuity, Titmus-fly and Lang tests were 89.3%, 74.2% and 86.1%, respectively. Sensitivity and specificity of the new test were 100% and 95.3%, respectively, which were not that different from those of the Randot preschool stereoacuity, Titmus-fly and Lang tests. We found that the digitalized, random-dot, stereoacuity test has a high success rate and can be appropriately used in medical examinations and follow-up tests for strabismus patients.


Assuntos
Pré-Escolar , Feminino , Humanos , Masculino , Distribuição por Idade , Estudo Comparativo , Gráficos por Computador , Percepção de Profundidade/fisiologia , Reprodutibilidade dos Testes , Sensibilidade e Especificidade , Software , Transtornos da Visão/diagnóstico , Testes Visuais/instrumentação , Visão Binocular