Chinese Journal of Medical Instrumentation ; (6): 263-265, 2019.
Artigo em Chinês | WPRIM | ID: wpr-772511

RESUMO

This study was to design a chronic obstructive pulmonary disease (COPD) screening equipment, based on the dual-differential pressure throttling technique. The technique combined a wide range, but low-resolution ratio sensor and a narrow range, but high-resolution ratio one. It can accurately detect the indexes of forced vital capacity (FVC), forced expiratory volume in one second (FEV), one second rate(FEV/FVC (%)), and achieve them in a low-cost way. The new designed machine will be compared with a British machine, named ML-3500. The correlations of FVC and FEV between new machine and ML-3500 were 0.998 and 0.999, respectively. The P values of paired test of these two indexes were over 0.05. Bland-Altman analysis of FVC, FEV and FEV/FVC (%) showed that more than 90% of the scatter points of the three parameters fell within the consistency interval. This machine can be used to accurately screen COPD and its low-cost would be advantage to promote in large population.


Assuntos
Humanos , Volume Expiratório Forçado , Doença Pulmonar Obstrutiva Crônica , Diagnóstico , Reprodutibilidade dos Testes , Testes de Função Respiratória , Economia , Volume de Ventilação Pulmonar , Capacidade Vital