Arq. bras. oftalmol ; 72(3): 346-350, May-June 2009. ilus, graf, tab
Artigo em Inglês | LILACS | ID: lil-521469

RESUMO

PURPOSE: To determine calibration errors of Goldmann applanation tonometers in ophthalmic clinics of Brasília, Brazil, and correlate the findings with variables related to tonometers model and utilization. METHODS: Tonometers from ophthalmic clinics in Brasília, Brazil, were checked for calibration errors. A standard Goldmann applanation tonometer checking tool was used to asses the calibration error. Only one trained individual made all verifications, with a masked reading of the results. Data on the model, age, daily use, frequency of calibration checking and the nature of the ophthalmic department - private or public - were collected and correlated with the observed errors. RESULTS: One hundred tonometers were checked for calibration. Forty seven percent (47/100) were out of 1 mmHg range at least at one point checking. Tonometers mounted to slit lamp, with less than 5 years, used in less than 20 patients daily, that had a calibration check on a yearly basis, and those from private office exhibit a lower rate of inaccuracy, but only the first variable was statistically significant. Sixty one percent of tonometers on public hospitals were out of calibration. CONCLUSION: Calibration of tonometers in the capital of Brazil is poor; those from general hospitals are worst, and this fact can lead to inaccurate detection and assessment of glaucoma patients, overall in the population under government assistance.


OBJETIVOS: Determinar os erros de calibração dos tonômetros de aplanação de Goldmann em clínicas oftalmológicas de Brasília, Brasil, e correlacioná-los a variáveis relativas ao modelo e à utilização dos aparelhos. MÉTODOS: Tonômetros de clínicas oftalmológicas de Brasília tiveram a calibragem aferida usando um cilindro padrão fornecido pelo fabricante dos aparelhos. Todas as aferições foram realizadas por um só examinador previamente treinado e a leitura das medidas foi mascarada por um observador independente. As medidas foram correlacionadas ao modelo, idade, utilização diária e frequência de aferição da calibração dos tonômetros, bem como o serviço em que o aparelho era utilizado - público ou privado. RESULTADOS: Cem tonômetros foram avaliados. Quarenta e sete por cento (47/100) apresentaram erros de leitura maiores ou iguais a 1 mmHg em pelo menos um ponto de aferição, sendo considerados descalibrados. Tonômetros fixos à lâmpada de fenda, com menos de 5 anos de fabricação, usados em menos de 20 pacientes por dia, que tinham a calibragem aferida ao menos anualmente e que eram utilizados em clínicas privadas apresentaram menos descalibragem, porém apenas a primeira variável foi estatisticamente significante. Sessenta e um por cento dos tonômetros de hospitais públicos estavam descalibrados. CONCLUSÕES: A calibragem de tonômetros avaliados em Brasília é precária. A situação é pior nos aparelhos dos hospitais públicos, podendo levar a erros na detecção e no acompanhamento de pacientes com glaucoma atendidos nesses serviços.


Assuntos
Humanos , Tonometria Ocular/instrumentação , Brasil , Estudos Transversais , Calibragem/normas , Análise de Falha de Equipamento/métodos , Hospitais Privados/estatística & dados numéricos , Hospitais Públicos/estatística & dados numéricos , Oftalmologia , Controle de Qualidade , Tonometria Ocular/normas
2.
Clinics ; 61(6): 511-514, 2006. graf, tab
Artigo em Inglês, Português | LILACS | ID: lil-439368

RESUMO

PURPOSE: The dynamic contour tonometer is a nonapplanation contact tonometer designed to be largely independent of the structural properties of the cornea. Theoretically, it may measure intraocular pressure most accurately in abnormally thinner corneas. This study compares intraocular pressure measurements by dynamic contour tonometry with Goldman applanation tonometry in eyes with normal corneas and eyes with advanced keratoconus. METHODS: A comparative case series. Subjects underwent intraocular pressure measurements by dynamic contour tonometry, Goldman applanation tonometry, ultrasonic pachymetry, and slit scanning topography. Exclusion criteria: any ocular pathology other than keratoconus, previous corneal or refractive surgery, stromal scarring due to acute hydrops, or any other corneal opacities. RESULTS: Ten patients with keratoconus were included in Group A, and 12 normal patients composed Group B according to the pre-established criteria. The mean Goldman tonometry measurement in group A was 10.3 ± 1.8 mm Hg and group B was 14.3 ± 0.75 mm Hg. (P = 0.024). In group A, the mean measurement with the dynamic contour tonometer was 14.6 ± 2.09 mm Hg, and in group B, it was 17.4 ± 3.1 mm Hg (P = 0.026). The difference between both methods of measurement in group A was statistically significant (P < 0.0002). CONCLUSION: Intraocular pressure readings with dynamic contour tonometry in the keratoconus group were significantly higher than Goldman measurements and lower when compared to the control group. As there are no published manometric studies in eyes with keratoconus, these lower intra-ocular readings with the dynamic contour tonometry could be related to the discrepancy between the radius of corneal curvature and its tip, the significant thinning of the cornea, or other corneal biomechanical abnormalities related to advanced keratoconus.


OBJETIVO: O tonômetro de contorno dinâmico é um aparelho de contato e não-aplanação projetado para ser independente das propriedades estruturais da córnea. Teoricamente, ele mede de forma mais precisa a pressão intra-ocular em córneas muito finas. Este estudo compara as medidas de pressão intra-ocular pelo tonômetro de contorno dinâmico com a tonometria de aplanação de Goldman em olhos normais e em olhos com ceratocone avançado. MÉTODOS: uma série comparativa de casos. Pacientes foram submetidos às medidas de pressão intra-ocular por ambos os métodos mencionados, paquimetria ultrassônica e topografia corneana de varredura. Critérios de exclusão: qualquer patologia ocular exceto ceratocone, cirurgia ceratorrefrativa prévia, opacidade estromal resultante de hidropsia aguda ou qualquer outra opacidade corneana. RESULTADOS: Dez pacientes com ceratocone foram incluídos no grupo A e 12 pacientes com córneas normais no grupo B conforme os critérios pré-estabelecidos. A média das medidas da tonometria de Goldman no grupo A foi de 10.3±1.8 mmHg e no grupo B de 14.3±0.75 mmHg. (p = 0,024). No grupo A, a média das medidas com o tonômetro de contorno dinâmico foi de 14.6±2,09 mmHg e, no grupo B,17,4±3,1 mmHg (p = 0,026). A diferença entre as medidas dos tonômetros no grupo A foi estatisticamente significante (p < 0.0002). CONCLUSÃO: As medidas da pressão intra-ocular com o tonômetro de contorno dinâmico nos pacientes com ceratocone foram significativamente maiores que as medidas do Goldman e menores quando comparadas com o grupo controle. Na ausência de estudos manométricos em olhos com ceratocone, estas baixas medidas com a tonometria de contorno dinâmico podem estar relacionadas com a discrepância entre o raio de curvatura da córnea e a ponta do tonômetro, os baixos valores paquimétricos, ou outras anormalidades da biomecânica corneana relacionadas com o ceratocone avançado.


Assuntos
Humanos , Masculino , Feminino , Adulto , Pressão Intraocular/fisiologia , Ceratocone/diagnóstico , Tonometria Ocular/métodos , Fenômenos Biomecânicos , Estudos de Casos e Controles , Topografia da Córnea , Ceratocone/fisiopatologia , Estudos Prospectivos , Reprodutibilidade dos Testes , Tonometria Ocular/normas