Your browser doesn't support javascript.
loading
Estudo comparativo entre as medidas da pressão ocular dos tonômetros de rebote, pneumático e de Goldmann / Comparative study between ocular pressure measurements made by rebot tonometer, pneumatic tonometer, Goldmann tonometry
Araujo, Hilkias Adachi; Araujo, Joao Pedro Vedovelli de; Medeiros, Felipe Beltrão de; Rebelo, Paulo Bernardo Ferreira; Medeiros, Hilton Arcoverde Gonçalves de.
  • Araujo, Hilkias Adachi; Adachi Oftalmologia. Macapá. BR
  • Araujo, Joao Pedro Vedovelli de; Adachi Oftalmologia. Macapá. BR
  • Medeiros, Felipe Beltrão de; Clinica de Olhos João Eugenio. Brasília. BR
  • Rebelo, Paulo Bernardo Ferreira; Clinica de Olhos João Eugenio. Brasília. BR
  • Medeiros, Hilton Arcoverde Gonçalves de; Clinica de Olhos João Eugenio. Brasília. BR
Rev. bras. oftalmol ; 78(3): 162-165, May-June 2019. tab, graf
Article in Portuguese | LILACS | ID: biblio-1013675
RESUMO
RESUMO

Objetivo:

Comparar o resultado entre dois tipos de tonômetros, tendo como padrão ouro a tonometria de aplanação, relacionando com a espessura corneana e suas validações como instrumentos de rastreio populacional.

Métodos:

Estudo transversal comparativo, realizado com 400 olhos do ambulatório de oftalmologia da Adachi Oftalmologia, em Macapá (Amapá), entre os valores médios da pressão ocular medidos pelo tonômetro Icare®, o Tonômetro Pneumático e o Tonômetro de Aplanação de Goldmann em função da espessura corneana. Foram divididos conforme a ECC em 3 grupos. Grupo 1 260 olhos com ECC< 530µ; grupo 2 217 olhos com ECC entre 530µ e 590µ; e, grupo 3 157 olhos com ECC>590µ.

Resultados:

A ECC média obtida na paquimetria foi de 557µ, variando de 651µ a 477µ (desvio padrão de 32.9). A PIO média obtida pela TP foi de 19,4 mmHg (com desvio padrão de 2,32); pelo Icare® foi de 16,7mmHg (desvio padrão de 2,12); e, na TAG foi de 15.5 mmHg (desvio padrão de 2,02). Para paquimetrias inferiores a 530µ e superiores a 590µ a PIOm não diferiu significativamente entre a TAG e Icare® (p = 0,232), sendo estatisticamente significativa diferença obtida para córneas com ECC entre 530 e 590 µ. A PIOm obtida pelo TP foi significativamente maior do que a pressão ocular média dos TAG e Icare® (p < 0,001 para ambos) em todos os grupos.

Conclusões:

Os três métodos podem ser utilizados em campanhas de rastreamento, porém o Icare® se mostrou mais confiável que a TP, com valores obtidos estatisticamente mais confiáveis quando comparados à TP. Existe suficiente concordância entre o Icare e a TAG para recomendar o seu uso.
ABSTRACT
ABSTRACT

Objective:

To compare the result between two types of tonometres, using as gold standard the aplanation tonometry, relating to the corneal thickness and its validations as instruments of population screening.

Methods:

A comparative cross-sectional study was carried out with 400 eyes from the ophthalmology clinic of Adachi Ophthalmology, in Macapá (Amapá), between the mean values of ocular pressure measured by the Icare® tonometro, the Pneumatic Tonometer and the Goldmann Flattening Tonometro as a function of corneal thickness. They were divided according to ECC in 3 groups. Group 1 260 eyes with ECC <530µ; group 2 217 eyes with ECC between 530µ and 590µ; and, group 3 157 eyes with ECC> 590µ.

Results:

The mean ECC obtained in pachymetry was 557µ, ranging from 651µ to 477µ (standard deviation of 32.9). The mean IOP obtained by PT was 19.4 mmHg (with a standard deviation of 2.32); by Icare® was 16.7mmHg (standard deviation of 2.12); and in the TAG it was 15.5 mmHg (standard deviation of 2.02). For pachymetry less than 530µ and greater than 590µ, PIOm did not differ significantly between TAG and Icare® (p = 0.232), and a statistically significant difference was obtained for corneas with ECC between 530 and 590 µ. The IOP obtained by the PT was significantly higher than the mean ocular pressure of the TAG and Icare® (p <0.001 for both) in all groups.

Conclusions:

All three methods can be used in screening campaigns, but Icare® proved to be more reliable than TP, with values obtained statistically more reliable when compared to TP. There is sufficient agreement between Icare and TAG to recommend its use.
Subject(s)


Full text: Available Index: LILACS (Americas) Main subject: Tonometry, Ocular / Glaucoma / Ocular Hypertension / Intraocular Pressure Type of study: Observational study / Prevalence study / Risk factors Limits: Adult / Female / Humans / Male Language: Portuguese Journal: Rev. bras. oftalmol Journal subject: Ophthalmology Year: 2019 Type: Article Affiliation country: Brazil Institution/Affiliation country: Adachi Oftalmologia/BR / Clinica de Olhos João Eugenio/BR

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Index: LILACS (Americas) Main subject: Tonometry, Ocular / Glaucoma / Ocular Hypertension / Intraocular Pressure Type of study: Observational study / Prevalence study / Risk factors Limits: Adult / Female / Humans / Male Language: Portuguese Journal: Rev. bras. oftalmol Journal subject: Ophthalmology Year: 2019 Type: Article Affiliation country: Brazil Institution/Affiliation country: Adachi Oftalmologia/BR / Clinica de Olhos João Eugenio/BR