Your browser doesn't support javascript.
loading
Adherencia y potenciales eventos adversos prevenidos durante la administración de medicamentos endovenosos empleando bombas de infusión inteligentes en cuatro unidades de cuidados intensivos en Colombia / Adherence and potential adverse events prevented during the administration of intravenous medications using Smart infusion pump in four intensive care units in Colombia
López-Romero, Luis Alberto; Cáceres-Rivera, Diana Isabel; Dueñas-Badillo, Ruby Emilse; Acosta-Barón, Jennifer Vanesa; Robles-Méndez, Ismery Yaneth; Rincón-Castillo, Diego; Parra de Rey, Gloria Isabel.
  • López-Romero, Luis Alberto; Fundación Cardiovascular de Colombia. Floridablanca. CO
  • Cáceres-Rivera, Diana Isabel; Fundación Cardiovascular de Colombia. Floridablanca. CO
  • Dueñas-Badillo, Ruby Emilse; Fundación Cardiovascular de Colombia. Floridablanca. CO
  • Acosta-Barón, Jennifer Vanesa; Fundación Cardiovascular de Colombia. Floridablanca. CO
  • Robles-Méndez, Ismery Yaneth; Fundación Cardiovascular de Colombia. Floridablanca. CO
  • Rincón-Castillo, Diego; Universidad Industrial de Santander. Bucaramanga. CO
  • Parra de Rey, Gloria Isabel; Fundación Cardiovascular de Colombia. Floridablanca. CO
Rev. Univ. Ind. Santander, Salud ; 51(4): 289-300, Septiembre 26, 2019. tab
Article in Spanish | LILACS | ID: biblio-1092259
RESUMEN
Resumen

Introducción:

Las bombas de infusión inteligentes, constituyen una herramienta útil para la administración segura de medicamentos endovenosos dado que permiten prevenir potenciales eventos adversos.

Objetivo:

Evaluar la adherencia y los potenciales eventos adversos prevenidos, durante la administración de medicamentos endovenosos empleando bombas de infusión inteligentes.

Metodología:

Estudio observacional, realizado en cuatro unidades de cuidados intensivos usando datos del software Hospira MedNetTM. Un análisis descriptivo fue llevado a cabo junto con un análisis bivariado empleando una prueba U de Mann-Whitney, una prueba de Kruskal-Wallis y un test de Bonferroni para evaluar la adherencia y los potenciales eventos adversos prevenidos por año y servicio.

Resultados:

La adherencia fue del 74,0%, se presentaron 78.299 alertas de seguridad y se previnieron 4,54% (n=16.288) potenciales eventos adversos. Se encontraron diferencias entre el primer y segundo año en la adherencia [Mediana 69,15 (Q164,2-Q375,5) Vs Mediana 84,2(Q172,15-Q389,05), p<0.001], adherencia a la seguridad [Mediana 87,1% (Q183,05-Q391,2) Vs Mediana 94,05 (Q189,95-Q396,2), p<0.001] y las ediciones de alertas de limite relativo [Mediana17,0 (Q18,5-Q324,5) Vs Mediana 12,0 (Q17,0-Q317,5), p=0.013]. La solución salina, la norepinefrina, el lactato de ringer, la piperacilina-tazobactam, la nitroglicerina y la heparina presentaron el mayor número de alertas de seguridad.

Conclusión:

Se encontró una buena adherencia (uso de la farmacoteca) y adherencia a la seguridad (indicador de uso de la bomba), con una reducción de los potenciales eventos adversos; así el uso de bombas inteligentes podría contribuir en la prevención de potenciales errores durante la administración de medicamentos endovenosos en la unidad de cuidados intensivos.
ABSTRACT
Abstract

Introduction:

Smart infusion pumps have become a useful tool for the safe administration of intravenous medications, since they allow the prevention of potential adverse events. Objetive To assess adherence and potential adverse events prevented during intravenous medication administration using smart infusion pumps.

Methods:

Observational study, conducted in four intensive care units using data from Hospira MedNetTM software. A descriptive analysis was carried out together with a bivariate analysis using a Mann-Whitney U test, a KruskalWallis test and a Bonferroni test to assess adherence and potential adverse events prevented by year and service.

Results:

Adherence was 74.0%, 78,299 safety alerts were presented and 4.54% (n = 16,288) potential adverse events were prevented. Differences were found between the first and second year in adherence [Median 69.15 (Q1 64.2-Q375.5) versus Median 84.2 (Q1 72.15-Q3 89.05), p<0.001]. Likewise safety adherence [Median 87.1% (Q1 83.05-Q3 91.2) versus Median 94.05 (Q1 89.95-Q3 96.2), p<0.001] and the relative limit alert editions [Median 17.0 (Q1 8.5-Q3 24.5) versus Median 12.0 (Q1 7.0-Q3 17.5), p=0.013]. The saline solution, norepinephrine, ringer's lactate, piperacillin-tazobactam, nitroglycerin and heparin presented the highest number of safety alerts.

Conclusions:

Adequate adherence (use of the drug library) and safety adherence (indicator of pump use) were found, with a reduction in potential adverse events; thus, the use of smart pumps could contribute to the prevention of potential errors during the administration of intravenous medications in the intensive care unit.
Subject(s)


Full text: Available Index: LILACS (Americas) Main subject: Medication Errors Type of study: Observational study Limits: Humans Country/Region as subject: South America / Colombia Language: Spanish Journal: Rev. Univ. Ind. Santander, Salud Journal subject: Medicine / Public Health Year: 2019 Type: Article Affiliation country: Colombia Institution/Affiliation country: Fundación Cardiovascular de Colombia/CO / Universidad Industrial de Santander/CO

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Index: LILACS (Americas) Main subject: Medication Errors Type of study: Observational study Limits: Humans Country/Region as subject: South America / Colombia Language: Spanish Journal: Rev. Univ. Ind. Santander, Salud Journal subject: Medicine / Public Health Year: 2019 Type: Article Affiliation country: Colombia Institution/Affiliation country: Fundación Cardiovascular de Colombia/CO / Universidad Industrial de Santander/CO