Your browser doesn't support javascript.
loading
Fatiga cíclica de cuatro sistemas reciprocantes en una canaleta dinámica de raíces simuladas / Cyclical fatigue of four reciprocating systems in a dynamic dump of simulating roots / Fadiga cíclica de quatro sistemas reciprocantes em uma com canais simulados
Correa Abad, Daicy Elizabeth; Hidalgo Araujo, Paola Daniela.
  • Correa Abad, Daicy Elizabeth; Universidad Central del Ecuador. Facultad de Odontología. Quito. EC
  • Hidalgo Araujo, Paola Daniela; Universidad Central del Ecuador. Facultad de Odontología. Quito. EC
Odontología (Ecuad.) ; 22(2): 45-59, 2020.
Article in Spanish | LILACS | ID: biblio-1103546
RESUMEN
En conductos radiculares curvos, la utilización de limas que eviten desvíos y fracturas del instrumento por fatiga y que mantengan curvaturas originales son aspectos importantes en el tratamiento endodóntico. Ob-jetivo Se evaluó la resistencia a la fatiga cíclica de cuatro sistemas reciprocantes en una canaleta dinámica que simuló tres curvaturas diferentes. Materiales y

Método:

Estudio in vitro; la muestra con 48 limas se dividió en cuatro grupos (n= 12/grupo) WaveOne® (WO), WaveOne® Gold (WOG), Reciproc® (R) y Reciproc® Blue (RB), cuatro limas de cada sistema se usaron en una canaleta que simuló tres angulaciones 45°, 60° y 90°; montadas en un motor X-Smart Plus®, las limas se introdujeron 23 mm en los conductos para ser activadas continuamente a 300/500 oscilaciones/minuto hasta que se produjo la fractura. Se registró tiempo y zona de fractura. Se realizó el test Kruskal Wallis con un nivel de confianza del 95% para las comparaciones. Resul-tados RB presentó mayor resistencia a la fatiga cíclica que WO y WOG en todas las angulaciones y que R a 90° (p= <0,05). R presentó mayor resistencia que WOG a 45° y 60°, sin embargo, a 90° WOG fue superior que R y WO (p<0,05). No hubo diferencias en cuanto a la zona de fractura que aconteció en la porción apical.

Conclusiones:

La resistencia se reduce cuando aumenta el grado de curvatura. RB fue significativamente más resistente a la fatiga cíclica que WO y WOG en todas las angulaciones y que R a 90°. La zona de fractura se produjo en la porción apical de las limas.
ABSTRACT
In the curved canals, the use of files that avoid deviations and fractures of the instrument due to fatigue and that maintain original curvatures are important aspects in endodontic treatment.

Objective:

The resistance to cyclic fatigue of four reciprocating systems was evaluated in a dynamic gutter that stimulated three different curvatures. Materials and

Method:

In vitro study, the sample with 48 files was divided into four groups (n = 12 /group) WaveOne® (WO), WaveOne® Gold (WOG), Reciproc® (R) and Reciproc® Blue (RB), four files from each system were used in a gutter that simulated three angulations 45o , 60 o and 90 o . The files which were mounted on an X-smart Plus® were inserted 23 mm into the conduits to be continuously activated at 300/500 oscillations per minute until the fracture occurred. Time and fracture zone were recorded. The Krustal Wallis test was performed with a 95% confidence level for the comparisons.

Results:

RB presented greater resis-tance to cyclic fatigue than WO and WOG at 45o and 60o, however, at 90o WOG was higher than R and WO (p<0.05). There were no differences regarding the fracture zone occurred in the apical portion.

Conclusions:

The resistance decreases when the degree of curvature increases. RB was significantly more resistance to cyclic fatigue than WO and WOG at all angulations and R at 90o. The fracture zone occurred in the apical portion of the files.
RESUMO
Em canais radiculares curvos, o uso de limas que evitem desvios e fraturas do instrumento devido à fadiga e que mantêm as curvaturas originais são aspectos importantes no tratamento endodôntico.

Objetivo:

avaliar a resistência à fadiga cíclica de quatro sistemas reciprocantes em uma canaleta artificial que simula três curva-turas diferentes. Materiais e

Método:

estudo in vitro; a amostra com 48 limas foi dividida em quatro grupos (n = 12 / grupo) WaveOne® (WO), WaveOne® Gold (WOG), Reciproc® (R) e Reciproc® Blue (RB), quatro limas de cada sistema foram usadas em uma canaleta que simulo três angulações 45 °, 60 ° e 90 °; se montaram em um motor X Smart Plus®, as limas foram inseridas 23 mm nos condutos para serem ativadas continuamente a 300/350 oscilações / minuto até ocorrer uma fratura. O tempo e a zona de fratura foram registrados. O teste de Kruskal Wallis foi realizado com um nível de confiança de 95% para as comparações.

Resultados:

O RB apresenta maior resistência à fadiga cíclica que o WO e o WOG em todas as angulações e que R a 90 ° (p = <0,05). R apresentou maior resistência que WOG a 45 ° e 60 °, no entanto, a 90 ° WOG foi maior que R e WO (p <0,05). No existiu diferença na zona de fratura que ocorreu na porção apical.

Conclusões:

a resistência diminui quando o grau de curvatura aumenta. RB foi significativamente mais resistente à fadiga cíclica do que WO e WOG em todos os ângulos e R a 90 °. A zona de fratura ocorre na parte apical das limas.
Subject(s)

Full text: Available Index: LILACS (Americas) Main subject: Dental Pulp Cavity / Endodontics / Flexural Strength Language: Spanish Journal: Odontología (Ecuad.) Journal subject: Ciˆncias da Sa£de / Odontologia Year: 2020 Type: Article Affiliation country: Ecuador Institution/Affiliation country: Universidad Central del Ecuador/EC

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: Dental Pulp Cavity / Endodontics / Flexural Strength Language: Spanish Journal: Odontología (Ecuad.) Journal subject: Ciˆncias da Sa£de / Odontologia Year: 2020 Type: Article Affiliation country: Ecuador Institution/Affiliation country: Universidad Central del Ecuador/EC